2.2.3. Safety instructions for the treadmill

 The treadmill belonging to the FDM-T measuring system is a very powerful device. 

For safe operation of the FDM-T system it is mandatory to adhere exactly to the 
safety regulations described in the following.

 The measuring procedure on the treadmill must never be commenced without a thorough 

instruction of the patient by trained personnel. No measurements may be taken without a supervisor.

 Do not place the treadmill on an unstable ground.

 Do not set up the system near a source of heating or in direct sunlight in front of a window as 

a strong rise in temperature can lead to inaccurate measuring results.

 Directly behind the treadmill it is mandatory for a safety zone of 2 m in length and 1 m in width

to be kept free, and ought to be padded (with a soft mat). No items may be left in this zone during 
operation (such as video camera, lighting equipment etc.).

 Dangerous drawing-in gaps are located at the rear end of the running belt and along its sides and

(if existing) on the elevating mechanism. Do not wear any loose clothing that could get caught up 
in the rollers. Make sure that if a person trips over, their long hair, loose clothing, jewellery, etc.,
do not encounter the rear part of the treadmill belt (e.g., wear a hair net). Due to danger of 
stumbling, do not place any clothing or jewellery on or within close proximity of the treadmill.

 Never use the treadmill without the safety clip being fastened to your clothing and always be sure 

that the folding mechanism (if existing) is properly locked during operation. (Please also be sure to 
read the safety instructions in the user manual supplied by the treadmill manufacturer.)

 During operation, the emergency STOP facilities must always be within easy reach for the user and

the operating personnel.

 The patient should walk slowly to begin with. Then gradually speed can be increased after a few 

minutes, depending on the patient’s physical condition. Improper or excessive strain through tests 
resp. measurements can have harmful effects on health.

 Never jump onto the running belt and never jump off it whilst it is running. Never stop walking 

while the treadmill is running, never turn round or step sideward or backwards. Should movement 
patterns of this kind be necessary for your measurements, please make sure to use a type of tread-
mill with a safety arch, protection against falling and a “fall stop” chest belt.

 Pull out the power plug before transporting the treadmill.

Содержание FDM-T-Sports

Страница 1: ...FDM T Sports Operating Instructions and Specifications...

Страница 2: ...edical GmbH Textrelease R 1 Stand 30 08 2021 REF Nr 79010094 Manufacturer zebris Medical GmbH Phone 49 0 7562 9726 300 Am Galgenbu hl 14 Fax 49 0 7562 9726 50 D 88316 Isny im Allg u Email sales zebris...

Страница 3: ...ensor 3 3 Technical Specifications FDM T measuring systems 3 4 Measuring principle FDM T system 3 5 Controls and connectors 3 6 LED indicator lights of the interface box 3 7 zebris SYNC 3 7 1 Synchron...

Страница 4: ...mendations for recording data 5 7 1 Treadmill analysis 5 7 2 Data recording 5 7 3 Walking speed 5 7 4 Posture 5 7 5 Weals 6 Maintenance and safety inspections 6 1 General maintenance procedures 6 2 Ma...

Страница 5: ...nsors and treadmill can also be used completely independently of each other and feature a separate power supply and CE mark The user manual for the FDM T measuring system therefore consists of several...

Страница 6: ...azard that can result in dam age of the device The notes explain the type of hazard and how it can be prevented This symbol indicates a potential hazard that can cause a loss of eyesight This warning...

Страница 7: ...dically used rooms or oxygen enriched atmospheres The devices must not be operated in proximity to e g engines or transformers with a high connected load as well as mains current lines as electrical o...

Страница 8: ...s delivered by zebris responsibility is assumed if Assembly extensions resettings changes or repairs were carried out through zebris or third parties authorised by zebris trained technicians or employ...

Страница 9: ...of fault or doubt If any fluids should penetrate the device it is mandatory for the device to undergo a technical safety test Damaged plug connections and leads are to be replaced by an authorized ser...

Страница 10: ...Make sure that if a person trips over their long hair loose clothing jewellery etc do not encounter the rear part of the treadmill belt e g wear a hair net Due to danger of stumbling do not place any...

Страница 11: ...plications using wheels are prohibited cycling wheelchair inline skating or roll ski as well as running shoes with spikes or studs Besides their extremely high risk of injury these can cause irreparab...

Страница 12: ...M software 3 2 Specifications FDM T sensor The sensors of the different FDM T systems only vary in size of the measuring area the number of single sensors included in the sensor module and the support...

Страница 13: ...sions L x W x H 160 x 80 4 x 136 cm folded 160 x 80 4 x 28 cm Step height 18 cm Incline adjustment manually 2 steps Max user weight 120 kg Colour white black Measuring range 1 120 N cm Sampling rate 1...

Страница 14: ...x 88 kg Dimensions L x W x H 177 x 69 x 115 cm Step height 16 cm Incline adjustment 0 to 15 in 1 steps Max user weight 120 kg Colour black green Measuring range 1 120 N cm Sampling rate 100 Hz Sensor...

Страница 15: ...16 cm Incline adjustment 0 to 15 in 1 steps Max user weight 150 kg Colour black green Measuring range 1 120 N cm 1 120 N cm Sampling rate 100 Hz 120 Hz optional 240 Hz Sensor area 94 8 x 40 6 cm 108...

Страница 16: ...L x W x H 188 x 70 x 142 cm Step height 20 cm Incline adjustment 3 bis 19 0 to 22 Max user weight 150 kg Colour white Measuring range 1 120 N cm Sampling rate 120 Hz Sensor area 91 x 39 cm Number of...

Страница 17: ...rted is determined for each individual sensor To do this the drive logic generates several sinus burst signals equivalent to the number of columns via the column decoder and transmits them to the resp...

Страница 18: ...to the interface box and connected to mains STANDBY Lights up if the power supply unit is connected to mains the USB socket is con nected to the PC and the hardware driver of the platform is installe...

Страница 19: ...to 5V 1 by an internal pullup resistor 2 7 k If this input is set to 0 V 0 i e by a switch or break contact than the SYNC IN is triggered Electrical spezifications Input Resistance Pull Up 5V 2 7 k V...

Страница 20: ...of the third party device either via start stop or sample by sample mode Electrical specifications Input Resistance Pull Up 5V 100 V VIH High Level Input Voltage 2 0 V VIL Low Level Input Voltage 0 8...

Страница 21: ...treadmill 21030071 USB2 0 cable A B Data connection between interface box and PC length 3 m 07200010 zebris FDM Software for operating system Windows 10 64 Bit 79010094 Hardware User Manual The print...

Страница 22: ...e or infrared includes software extension 01540194 SYNCLightCam 30 Hz Combined solution with Camera and illumination 5 m USB2 0 Cable synchronization cable inclusive software extension 01831105 SYNCLi...

Страница 23: ...Modul virtual training Software extension interactive treadmill training on a virtual forest path contains levels with different levels of difficulty and an additional level Acoustic Cueing 01530100 O...

Страница 24: ...ta the camera must be connected to the galvanically isolated Viodeosync socket on the interface box Videosync and USB socket at the backside of the tradmill PDM C sensor 4 1 1 Connection to the zebris...

Страница 25: ...r supply Mascot 0813105 Power supply Mascot 9926 SYNCLightCam 1 21030071 USB cable A B 21030320 SYNCCam USB cable A B 01830041 Videosync control cable 2 5 21030312 Videosync control extension 01830041...

Страница 26: ...sure distribution measurement All adjustments of the camera are carried out via hardware setup integrated to the zebris FDM Software The camera is connected to the PC by a USB cable of type A B includ...

Страница 27: ...nt All adjustments of the camera are carried out via hardware setup integrated to the zebris FDM software The camera is connected to the PC by a USB cable of type A C included within the shipment The...

Страница 28: ...ization with special synchronization cable included in the scope of delivery Infrared Synchronization via infrared signal without synchronization cable not compatible with every zebris pressure plate...

Страница 29: ...a USB cable of type A B included within the shipment The SYNCLightCam is equipped with inch tripod threads and can be adapted to zebris tripods as wells as commercially available camera tripods Furthe...

Страница 30: ...DM T Sports Technical specifications REF 01540194 Variant 30 Hz Dimensions 220 x 183 x 80 mm B x W x T Weight ca 790 g Power supply 24 V 36 W Resolution 1920 x 1080 Pixel Full HD Autofocus Frame Rate...

Страница 31: ...which results in reduced brightness to prevent the LEDs from being damaged by excessive heat Power supply unit For operation of the SYNCLight plus a power supply unit needs to be connected REF 331022...

Страница 32: ...nd 1 m in width must be kept free directly behind the treadmill No items may be left in this zone video camera etc For setup installation and safety related instructions of the treadmill please refer...

Страница 33: ...acteristics Connect only if they are compatible For operating the FDM T sensor exclusively use the power adapter approved by zebris which is suitable for the power supply of all FDM T systems Seite 33...

Страница 34: ...of the treadmill and the FDM T power supply unit must be done at a separate wall socket It is not permissible to use extension cables and or multiple sockets Seite 34 69 Seite 34 41 FDM T Sports 5 2...

Страница 35: ...tuation 5 7 1 Treadmill analysis Walking on the treadmill is unfamiliar to most people and takes a certain time to get used to it It is therefore of advantage to familiarize the test person with this...

Страница 36: ...e sliding surface underneath the treadmill belt is no longer sufficiently lubricated g A malfunction or a defect is suspected or has been detected Check regularly approx every 25 operating hours wheth...

Страница 37: ...s Before each use of the measuring system it is recommended for safety reasons to check the correct state of all the connection cables as well as the mains cable mains plug and mains socket Should cer...

Страница 38: ...e shown for a condition without any load In addition the pressure distribution images are to be checked regularly for untypical measuring patterns These include above all line or column shaped measuri...

Страница 39: ...d to the PC and the device driver is correctly installed Are all the other components of the measuring system infrared synchronization with zebris video camera connected correctly Check list for notin...

Страница 40: ...disinfectant s recommended duration of impact Please make sure to switch off the device and pull the mains plug out of the socket before you commence disinfecting and cleaning Do not use any aggressiv...

Страница 41: ...e the measuring system can be returned to zebris Medical GmbH at the end of its service life at the customer s own expense and will be forwarded to special recycling companies without any further cost...

Отзывы: