zebris Medical GmbH
FDM Technical Data and Operating Instructions
Page 46/54
6.4
Cleaning and disinfection
6.4.1 Cleaning
The platform and accessories are cleaned with a moist cloth while the device is switched off
and the mains plug taken out.
NOTE
Do not use any aggressive agents to clean the measuring system.
WARNING
Please be absolutely sure to switch off the device and pull the mains
plug out of the socket before you commence disinfecting and cleaning
6.4.2 Manuelle Desinfektion
The platform can be disinfected by wiping over with suitable agents. To clean, wipe the
platform with a cloth soaked in disinfection liquid.
WARNING
No spray disinfection
!
Spray disinfection can destroy the highly precise
measurement sensors of the platform.
Recommended disinfection agent
Composition approx. 25% ethanol, 35% Propanol
E.g. Mikrozid Liquid / Schülke & Mayr or similar agents
NOTE
If you apply disinfection agent be sure to follow the recommendations
given by the manufacturer of the disinfection agent strictly. Especially
consider the rules concerning the commended application time of the
agent.
WARNING
Due to danger of confusion, chemicals that are necessary for the dis-
infection or cleansing exclusively must be stored, prepared and pro-
vided in containers that are appropriate for this purpose.
NOTE
In order to demonstrate that disinfection was successfully done, it is
recommended to put up a sign on the platform, saying “disinfected”.