1.8 Measuring principle
The system contains a measuring matrix consisting of capacitive pressure sensors that
are arranged in columns and lines running closely next to each other. For determining
the force distribution over the measuring matrix the capacity proportional to the force
exerted is determined for each individual sensor. To do this, the drive logic generates a
number of sinus burst signals equivalent to the number of columns via the column
decoder
,
and transmits them to the respective measuring column. The analog signal
coupled into the shift register over the lines is proportional to the pressure-dependent
capacity and is passed on for further processing to the control and signal-processing
electronics and transmitted to the PC from there and shown on the display.