Setting Alarm Out Status
Alarm outputs can be configured to enable or disable
which alarms deactivate the alarm output contact
located on TB1, terminals #17 and #18,
Wiring Control Document on page 116 in Appendix D.
Only an entire alarm area may be disabled for example,
pump alarms.
CA
CA
CA
CA
CAUTION:
UTION:
UTION:
UTION:
UTION:
Alarm outputs are critical for monitoring system perfor-
mance. Outputs should not be disabled except for
testing/troubleshooting.
Choose *Set in the main menu, figure 149.
Choose *Par – parameters in the Set Selection menu,
figure 150.
Choose *Flow or *Time in the Set Parameters menu,
figures 151.
Choose *Alarm from the Proportional-to-Time or
Proportional-to-Flow menu, figure 152.
Choose *Set from the Alarm Out Status menu, figure
153.
To set the pump alarm status, figure 154, press and
release the Select key. The entered value will flash
when it is chosen. Use the Down Arrow or Up Arrow
keys to change the setting. Press the Enter key to load
the entry into memory. The display will stop flashing
when the entered value is loaded into the memory.
Press the Down Arrow key to advance to the next
parameter.
To set the battery alarm status, figure 155, press and
release the Select key. The entered value will begin to
flash when chosen. Use the Up Arrow or Down Arrow
keys to change the value. When a new value has been
chosen, press the Enter key to store the new value into
memory. The entered value will stop flashing when it
has been loaded into memory. Press the Down Arrow
key to advance to the next parameter.
SSSSS
ECTION
ECTION
ECTION
ECTION
ECTION
9: S
9: S
9: S
9: S
9: S
ETTING
ETTING
ETTING
ETTING
ETTING
& T
& T
& T
& T
& T
ESTING
ESTING
ESTING
ESTING
ESTING
A
A
A
A
A
LARMS
LARMS
LARMS
LARMS
LARMS
Or
pmp bat vmtr sig tnk
Proportional to Flow
*Set *Alarm * Esc
pmp bat vmtr sig tnk
Proportional to Time
*Set *Alarm * Esc
Figure 149
pmp bat vmtr sig tnk
PropFlow Idle
*Strt *Dsp *Set
pmp bat vmtr sig tnk
Set Selection
*Par *Cal *Esc
pmp bat vmtr sig tnk
Set Parameters
*Flow *Time *Esc
pmp bat vmtr sig tnk
Alarm Output Status
*Set *Test *Esc
Select
Enter
pmp bat vmtr sig tnk
Pump Alarm
Enabled
pmp bat vmtr sig tnk
Pump Alarm
Disabled
pmp bat vmtr sig tnk
Battery Alarm
Enabled
pmp bat vmtr sig tnk
Battery Alarm
Disabled
Select
Enter
Figure 150
Figure 151
Figure 152
Figure 153
Figure 154
Figure 155