Oxygen Concentrator 

Model

7F-3 / 7F-5B 

User’s Manual 

 

 

 

 

 

 

DO NOT OPERATE THIS UNIT WITHOUT FIRST 

  READING AND UNDERSTANDING THIS MANUAL 

Содержание 7F-3

Страница 1: ...Oxygen Concentrator Model 7F 3 7F 5B User s Manual DO NOT OPERATE THIS UNIT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL ...

Страница 2: ... 1 TABLE OF CONTENTS SAFETY NOTES 2 6 FEATURES 7 9 HANDLING 10 OPERATING INSTALLATION 11 16 MAINTENANCE 17 18 TROUBLESHOOTING 19 20 OTHER ATTENTION ITEMS 21 22 EMC INFORMATION 23 25 ...

Страница 3: ...ygen poisoning and oxygen allergy user patient DO NOT using this oxygen concentrator SIDE EFFECT No side effect Symbol Description Describes a hazard or unsafe practice that if not avoided can result in severe bodily injury death or property damage Describes a hazard or unsafe practice that if not avoided can result in minor bodily injury or property damage Ⅰ IMPORTANT MESSAGE To reduce the risk o...

Страница 4: ... displayed Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air Failure to observe this warning can result in severe fire property damage and cause physical injury or death It can take up to 10 minutes to stabilize to deliver the set flowrate and concentration of oxygen to user after switching on the oxygen concentrator F...

Страница 5: ... must be located in another room at least 2 5m from the bath DO NOT come in contact with the oxygen concentrator while wet DO NOT place or store oxygen concentrator where it can drop into water or other liquid DO NOT reach for oxygen concentrator that has fallen into water UNPLUG IT IMMEDIATELY Oxygen concentrator should NEVER be left unattended when plugged in Oxygen concentrator is to be used on...

Страница 6: ...ents excepts single use consumable materials And the Service Life is equal to the shelf life Be aware that the power cord and or tubing could present a tripping or strangulation hazard Always place the power cord and or tubing in a manner that prevents crushing by casters or others DO NOT modify the oxygen concentrator in any way Modifications could result in hazards to the user There is a risk of...

Страница 7: ...emergency while undergoing oxygen therapy seek medical assistance immediately to avoid harm Geriatric paediatric or any other patient unable to communicate discomfort can require additional monitoring and or a distributed alarm system to convey the information about the discomfort and or the medical urgency to the responsible care giver to avoid harm Smoking during oxygen therapy is dangerous and ...

Страница 8: ...reen Timing turns off function using convenience Compressor pressure relief valve help the device safer Power interruption alarm function Device failure alarm function including pressure failure compressor failure low oxygen concentration Compressor with over heat protect function to gain in safety of the compressor and the concentrator Ⅲ SPECIFICATIONS 1 Power Supply 230V 50Hz 2 Input Power 7F 3 ...

Страница 9: ...ow L min Oxygen Concentration 0 5 95 5 87 1 95 5 87 2 95 5 87 3 95 5 87 4 95 5 87 5 95 5 87 9 Altitude Not higher than 2000 meters above sea level 10 Net Weight 24kg Dimension 47 28 5 55 6 cm 11 Work system work continuously 12 Minimum Operating Time 30minutes 13 Electric classification class II equipment type BF applied part IP21 14 Electric classification over voltage category II pollution degre...

Страница 10: ...l Cannula 41 in normal condition Applied part Temperature Nasal Cannula 43 in single fault condition 18 Cannula length DO NOT more than 15 2m and no twist 19 Storage and Transportation Condition Temperature range 20 60 Relative humidity 10 93 Non condensing Atmosphere pressure 700hPa 1060hPa The device should be stored with no strong sunlight no corrosive gas and well ventilated indoor area The de...

Страница 11: ... damage is evident please notify the carrier or local dealer 2 Remove all loose packing from the carton 3 Carefully remove all the components from the carton Ⅱ INSPECTION 1 Examine exterior of the oxygen concentrator for nicks dents scratches or other damages 2 Inspect all components Ⅲ STORAGE 1 Store the repackaged oxygen concentrator in a dry area 2 DO NOT place other objects on top of the repac...

Страница 12: ... 11 OPERATING INSTALLATION Ⅰ FEATURE VIEW Figure 1 view of the machine view of the display panel Figure 2 Humidifier Components ...

Страница 13: ...or plug if it is not working properly if it has been dropped or damaged or dropped into liquid call Qualified Service Personnel for examination and repair 2 Keep the cord away from HEATED or HOT surfaces 3 Do not move or relocate concentrator by pulling on the cord 4 Do not use extension cords with this unit NOTE Concentrator may be used during the initial start warm up time approximately 30 min w...

Страница 14: ...sh an oxygen delivery tube fire and stop the flow of oxygen if the tube is accidentally ignited And the firesafe valve is flow direction sensitive component the install direction must correct Note The firesafe valve is maintenance free and has an intended life of 4 years Dispose of any firesafe valve 4 years after the manufacturing date state on the device label Once actuated the firesafe valve ca...

Страница 15: ... concentration values The explanations of the indicator light functions are as follows Explanation of the indicator light and alarm Symbol Status Indicator lights alarm I O SYSTEM OKAY O2 concentration is greater than 82 Green O2 concentration is less than 82 Yellow Alarm SYSTEM FAILURE Pressure failure alarm Or Compressor circuit failure alarm Or over temperature alarm low flowrate Yellow Alarm P...

Страница 16: ...r immediately 7 Power loss alarm alarm sound no display no machine working Please check the power input NOTE Concentrator will reach to most stable state after warm up approximately 30 minutes Ⅴ TIME SETTING User can press the timing button to setting the operating time in 1 99 minutes When it start working the display shows elapsed time and setting time it means that timing function is closed It ...

Страница 17: ...tion source and smoking prohibited No smoking Keep Up Fragile Temperature limit Humidity limitation Atmospheric pressure limitation Refer to instruction manual Keep Dry Stacking Limitation Manufacturer European Representative IP21 Enclosures Protection Classification The first characteristic number 2 Protected against access to hazardous parts with a finger The second characteristic number 1 Prote...

Страница 18: ... cabinet Clean the cabinet with a mild household cleaner and non abrasive cloth or sponge at least one time a month Do Not drop any liquid into the seam of the device Ⅱ CLEAN OR REPLACE FILTER Please clean or replace the filters in time it s very important to protect compressor and to extend the device life 1 Clean the I stage air inlet once a month Figure 6 2 The II stage filter mesh cleans once ...

Страница 19: ...e a week at least or as necessary 4 Replace clean water into humidifier every day before using Note During both normal condition and single fault condition the cabinet humidifier and nasal cannula can become contaminated with body fluids or expired gases To reduce the infection risk please perform the maintenance routinely 4 REPLACE FUSE TUBE Fuse tube for net power Figure 10 Type T5AH250V Φ5 20 C...

Страница 20: ...ty valve 3 The connection between oxygen output and humidifier is leaking 3 Re install the humidifier tubing 4 The accessory nasal cannual mask humidifier tubing etc is leaking 4 Replace the leaked accessory 5 If the phenomenon is still appeared please contact the supplier Concentrator is working but yellow light illuminate alarm 1 Oxygen Concentration 82 1 Clean or replace the filter 2 Oxygen flo...

Страница 21: ...s Remove the vapor or liquid drops method Turn on the device and connect the tubing to the oxygen outlet Use your finger to block and release the end of the tubing several times to remove the moisture and liquid 1 Poor ventilation around the around the device so operating temperature is too high 1 Make sure the device is at least 10 cm away from the walls other jam or heater 2 The water which adde...

Страница 22: ...or Heat Exchanger Valve Assembly Molecular Sieve Oxygen Nitrogen Separation Oxygen Nitrogen Commute Valve Muffler Product Tank Bacilli Filter Flowmeter Check Valve Oxygen Outlet Humidifier Pressure Regulator Transformer Display Board Main Board Assemb ValveⅠ Assemb ValveⅡ Compressor Fan Oxygen Sensor ...

Страница 23: ...ical Manual 1 piece 3 Fuse tube T5AH250V φ5 20 2pcs 4 Firesafe Cannula Valve Bpr firesafe 1 piece 5 connect tube φ5 φ9 40mm silicone tube 6 Nasal Oxygen Cannula 2m 1 piece Ⅳ DEVICE DISPOSAL Contact the local cit or town offices for proper disposal instructions for oxygen concentrator ...

Страница 24: ...c immunity of this equipment and result in improper operation Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30cm 12 inches to any part of the Oxygen Concentrator including cables specified by the manufacturer Otherwise degradation of the performance of this equipment could result Electric and magnetic environme...

Страница 25: ...ariations on power supply input lines IEC 61000 4 11 0 UT 0 5 cycle at 0 45 90 135 180 225 270 and 315 0 UT 1 cycle 70 UT 25 30 cycles at 0 Voltage short interruptions 0 UT 250 300 cycles at 0 Power frequency 50 Hz magnetic IEC 61000 4 8 30 A m Radiated RF EM fields IEC 61000 4 3 10V m 80MHz 2 7GHz 80 AM at 1kHz Conducted disturbances induced by RF fields IEC 61000 4 6 3V m 0 15MHz 80MHz 6V in ISM...

Страница 26: ...80 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation 18 Hz 2 0 3 28 870 930 1720 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation 217 Hz 2 0 3 28 1845 1970 2450 2400 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulation 217 Hz 2 0 3 28 5240 5100 5800 WLAN 802 11 a n Pulse modulation 217 Hz 0 2 0 3 9 5500 5785 NOTE If ...

Страница 27: ...nyang Jiangsu P R China TEL 86 511 86900833 Shanghai International Holding Corp GmbH Europe Address Eiffestrasse 80 20537 Hamburg Germany TEL 0049 40 2513175 Notified Body Name TÜV SÜD Product Service GmbH Address Ridlerstr 65 80339 MÜnchen Germany Document No YY 7F 3 5B 01 B 1 2019 06 ...