background image

GATE

adjusting the height of the gate.Only when the waveform is higher

than the gate, the gauge can take the echo and show the value. Attention: this
will only show when the GATE specification is highlighted. The first
intersection point between the waveform and gate is pointed by a red arrow,
which can help judging whether the thickness value is correct (the red arrow
should point the front of the first bottom echo if correctly tested).

3.5 Real Cases Analysis of A-Scan

1

. When measuring the thickness, it’s possible that that over small gains

prevent the precise results. As showed in the following figure, the thickness of
the testing object is about 5mm, but as for the very small gain, the measuring
result is 10.77mm as the first echo has not broken the gate and the gate locates
the second echo automatically. This is obviously an incorrect result, and
customer can pull up the echo by enhancing the gain setting to make the first
echo brake through the gate to achieve the correct measurement.

3.12 REAL CASE 1

2

. There are some defects in the testing object, and the gate locks the defect

echoes. As showed in the following figure, the thickness of the testing object
is about 10mm, but as there are obvious defects (the defect echoes are showed
on the display) and the gate locks the defect echoes which have broken the
gate, thus, the testing result shown is the thickness of the defect area. The right
measurement can be realized by adjusting the gate setting above the defect
echoes.

3.13 REAL CASE 2

3

. If there are some surface faults in the testing object and the gate locks the

defect echoes, the measuring result will be the thickness of the defect area. In
this condition, the customer can use the BLANK function to get the correct
testing result. As showed in the following figure, the line on the bottom of the
screen, which we use to shield the defect echoes, indicates the blank confines
and makes the gate not catch the echoes within the blank confines, thus, the
right thickness value is acquired.

3.14 REAL CASE 3

17

18

3.11 waveform of the 0.02in sheet by PT04

Содержание UM-5 Series

Страница 1: ...15 3 5 Real Cases Analysis of A Scan 17 3 6 Operation of B scan Interface 19 3 6 1 B scan Display 19 3 6 2 Introduction of B Scan 19 3 7 Dual echo Thru coating Mode 20 3 7 1 A scan Interface in Dual...

Страница 2: ...splay is available by showing the echoes which enables the more extensive measure range more precise result and very little measurement error and finally makes the measuring processes more controllabl...

Страница 3: ...LF FULL Display Mode Thickness Value Min Max capture DIFF RR A Scan B Scan V Path Correction Automatic compensate the non linearity of dual element probe Update Rate Selectable 4Hz 8Hz 16Hz Material V...

Страница 4: ...eat capacity data storage function Stores 100 000 thickness values 1000 waveforms only UM 5DL 15 Measure the net thickness of the workpiece through the coating layer only UM 5D UM 5DL 2 Keyboard Funct...

Страница 5: ...If the material velocity is known for example the velocity of steel is 5900m s The sound velocity can be setting manually 3 1 1 Probe zero calibration Note The calibration result is 4 00mm only under...

Страница 6: ...on adjusting options like FILE NUMBER MEASURING MODE VIEW MODE PROBE SETUP MINIMUM ALARM MAXIMUM ALARM NORMAL THICKNESS THE MINIMUM OF B SCAN THE MAXIMUM OF B SCAN RECTIFICATION RECTIFICATION WAVEFORM...

Страница 7: ...TIFICATION MODE RF full wave half half RF RF describes the complete echo waveform full wave indicates the half echo and the overturned half echo half means putting off the half echo and turn the half...

Страница 8: ...erence between measured and nominal thickness and reduction difference as a percentage of nominal thickness as well as the current measured and nominal thickness values Before using this mode presetti...

Страница 9: ...mall increments while left and right for large increments GAIN Adjust the amplification of echo signal by instrument to allow manual increase or decrease in ldB This function is very useful for the te...

Страница 10: ...echo brake through the gate to achieve the correct measurement 3 12 REAL CASE 1 2 There are some defects in the testing object and the gate locks the defect echoes As showed in the following figure th...

Страница 11: ...ng or paint layer on the surface of the workpiece it will bring considerable error to the measurement results UM 5D and UM 5DL have the echo echo measurement method which can accurately measure the ac...

Страница 12: ...kness value 7 delete the selected data 8 The current cell selected Whether it is in the interface of thickness value A scan or B scan the current measured thickness values will be stored for short pre...

Страница 13: ...t 5 2 Measuring Methods 1 SINGLE POINT MEASUREMENT Using the probe to measure a random point on the surface of the object the reading displayed is the thickness value 2 DOUBLE POINT MEASUREMENT Measur...

Страница 14: ...time the gauge will not be turned on after running out of power 6 2Considerations 1 Please be cautious of the zero block s getting rust as couplant will be spread on the surface of it when calibrating...

Страница 15: ...5800 Mercury 0 057 1400 Molybdenum 0 250 6300 Monel 0 210 5400 Neoprene 0 063 1600 Nickel 0 220 5600 Nylon 6 6 0 100 2600 Oil SAE 30 0 067 1700 Platinum 0 130 3300 Plexiglass 0 110 1700 Polyethylene...

Отзывы: