background image

10

11

window will have to be cleaned off.  After installing the o-ring, ensure it is 

clean.  If necessary, wipe clean with a lens tissue.

Locate the o-ring lubricant included with the new sensor cap.  Apply a thin 

6. 

coat  of  o-ring  lubricant  to  the  installed  o-ring.    After  application,  there 

should be a thin coat of o-ring lubricant on the o-ring only.  Remove any 

excess o-ring lubricant from the o-ring and/or probe with a lens tissue. 

Clean the clear surface of the optical DO sensor window (figure 1) with a 

7. 

lens cleaning tissue.

Remove the new sensor cap from its hydrated container and dry the inside 

8. 

cavity  of  the  sensor  cap  with  lens  tissue.    Make  sure  that  the  cavity  is 

completely dry and clean before proceeding with the installation.  Using a 

clockwise motion, thread the new sensor cap onto the probe assembly until 

it is finger-tight.  The o-ring should be compressed between the sensor cap 

and probe.  Do not over-tighten the sensor cap and do not use any tools for 

the installation process.

Clean the stir paddle and then reinstall it on the probe by pushing it straight 

9. 

into place.    

Store the sensor in a BOD bottle with a small amount of water (approximately 

10. 

40 ml).

Follow the procedures below for configuring the ProODO instrument for 

11. 

the new Sensor Cap’s coefficients.

Configuring ProODO Instrument for new Sensor Cap
After  installing  a  new  Sensor  Cap,  connect  the  probe/cable  assembly  to  the 

ProODO instrument and turn the instrument on.  Locate the Calibration Code 

label at the top of the Sensor Cap instruction sheet and note the six numbers 

which  are  listed  as  k1  through  k5  and  kC.    These  six  numbers  contain  the 

calibration code specific to the sensor cap that was just installed.  Follow the 

instructions below to enter the new calibration coefficients into your ProODO 

instrument.

Press the 

1. 

Probe

 

 on the ProODO instrument, highlight

 DO

, and press 

enter.   

Highlight 

2. 

Sensor Cap

 and press enter.

Highlight 

3. 

Sensor Cap Coefficients

 and press enter.

k1 will be highlighted.  Press enter to access 

4. 

the  numeric  entry  screen.    Enter  the  new  k1 

coefficient  from  the  Calibration  Code  label.  

After  the  Calibration  Code  has  been  entered 

correctly, press enter to confirm.  

Highlight k2 through kC in turn and use the 

5. 

numeric entry screen to enter the corresponding 

new coefficient from the Calibration Code label 

as described in step 4.  Press enter after each entry and then proceed to the 

next k selection.

After  all  the  new  coefficients  have  been  entered,  highlight 

6. 

Update 

Coefficients

 and press enter.  

A  message  will  appear  warning  that  you  will  be  overwriting  the  current 

7. 

sensor cap coefficients and you should confirm that you wish to carry out 

this action.  Highlight 

Yes

 and press enter to confirm the new coefficients.

After updating the Coefficients, the Serial # in the Sensor Cap menu will be 

updated automatically based on your entries.  

If  an  error  is  made  in  entering  the  Sensor  Cap  Coefficients,  the  instrument 

will block the update and an error message will appear on the display.  If you 

see  this  error  message,  re-enter  the  coefficients  checking  them  carefully  for 

correct transcription from the Calibration Code label prior to selecting 

Update 

Coefficients

.      If  you  continue  to  get  an  error  message  after  several  entry 

attempts, contact YSI Technical Support for assistance.   

After  entering  the  new  Sensor  Cap  coefficients,  perform  a  1-point  DO 

calibration. 

REHYDRATING THE SENSOR CAP  

The Sensor Cap must remain in a moist environment for proper operation; see 

Sensor Storage

 

for storage recommendations.  If you inadvertently leave your 

sensor exposed to ambient air for a period of more than approximately 8 hours 

it may dry out.  If the sensor cap is allowed to dry out, it is likely to drift slightly 

at the beginning of your next study unless it is rehydrated.  If the cap dries out, 

you can rehydrate it by soaking the probe tip with the sensor cap installed in 

warm (room temperature) tap water for 24 hours.  After rehydration is complete, 

perform  a  1-point  DO  calibration  and  be  sure  to  store  the  probe  in  a  moist 

environment.   

sensOr maintenance - temperatUre

You must keep the temperature portion of the sensor free of build up.  Other 

than that, the sensor requires no maintenance.  A soft bristle brush can be used 

to scrub the temperature sensor if needed.  While cleaning, be sure to only scrub 

the temperature sensor and not the sensor cap.  The sensor cap will be damaged 

if cleaned with anything abrasive.    

Содержание WE KNOW ProOBOD

Страница 1: ...USER MANUAL ...

Страница 2: ...ntenance and Storage 8 Updating Instrument and Probe Firmware 8 Sensor Maintenance Dissolved Oxygen 8 Sensor Maintenance Temperature 11 Sensor Storage 12 Troubleshooting 12 Dissolved Oxygen Readings 12 Help 13 Error Status messages 13 Restore Default Calibration Values 14 Probe Specifications 15 Accessories Parts List 16 Declaration of Conformity 17 Recycling 19 Contact Information 20 Ordering and...

Страница 3: ...elected by YSI Repair or replacement will be made and the product returned transportation prepaid Repaired or replaced products are warranted for the balance of the original warranty period or at least 90 days from date of repair or replacement LIMITATION OF WARRANTY This Warranty does not apply to any YSI product damage or failure caused by Failure to install operate or use the product in accorda...

Страница 4: ...tional outlet adapters are also included The ProOBOD sensor measures dissolved oxygen in water using lifetime luminescence technology and uses a digital signal to send information between the instrument and probe The probe features a stirrer and motor to aid in sample movement to keep solids from settling at the bottom of the sample bottle and to speed up the response of the sensor Key advantages ...

Страница 5: ...nstruction sheet is important because it contains the calibration coefficients for your sensor cap After using this sheet for general probe setup be sure to store it in a safe place in case you need to reload these calibration coefficients in the unlikely event that they are ever deleted from the instrument and probe Note A new cable probe assembly already has a sensor cap installed and the sensor...

Страница 6: ...r cap needs replaced See the Sensor Maintenance Dissolved Oxygen section of this manual or the instruction sheet included with the replacement sensor cap for instructions on updating the Sensor Cap Coefficients When updating the Coefficients the sensor cap serial will be updated automatically based on your entries The Temperature Coefficient should not be modified unless instructed by YSI Technica...

Страница 7: ...nd then press Cal to complete the calibration The message line at the bottom of the screen will display Calibrating Channel followed by Calibration Successful Press Esc to cancel the calibration i Temperature All ProOBOD probes have built in temperature sensors Temperature calibration is not required nor is it available To set the units press the Probe key on the ProODO instrument and select Displ...

Страница 8: ...en replacing a Sensor Cap the Sensor Cap Coefficients must be manually updated in the ProODO instrument The instruction sheet shipped with the replacement OBOD sensor cap includes the calibration coefficients that are specific to your Sensor Cap u IMPORTANT Be sure to save the OBOD Sensor Cap instruction sheet in case you need to reload the calibration coefficients The replacement OBOD Sensor Cap ...

Страница 9: ...Calibration Code has been entered correctly press enter to confirm Highlight K2 through KC in turn and use the 5 numeric entry screen to enter the corresponding new coefficient from the Calibration Code label as described in step 4 Press enter after each entry and then proceed to the next K selection After all the new coefficients have been entered highlight 6 Update Coefficients and press enter A...

Страница 10: ... sensor cap and then recalibrate If you are still getting erroneous dissolved oxygen readings try replacing the sensor cap and then recalibrate If the erroneous readings continue contact YSI Technical Support to help determine the next step Inspecting the Sensor Cap for Damage Caution Avoid touching the sensing end of the sensor cap during the following maintenance procedures If DO measurements se...

Страница 11: ... Values Occasionally the instrument may need to have the factory calibration default values restored To perform a factory reset press Calibrate highlight Restore Default Cal and press enter Select the parameter you wish to restore either DO or Barometer and press enter After selecting barometer or DO you will be asked to confirm the operation Highlight Yes and press enter P robe S pecifications Th...

Отзывы: