The ProSample P and ProSample PM can be configured to sample based on measurements/signals from analog and
digital I/O devices, while the ProSample P-12 and ProSample PM-12 can additionally be configured to sample based
on measurements from SDI-12 devices (
e.g.
EXO3 sonde).
Flow Signal Cable Wiring
The flow signal cable (item # 630147) can be used to send a digital, analog, or event signal to the ProSample. One
end of this cable has flying leads so it can be connected to an analog and/or digital device, while the other end can be
connected to the Signal Input on the ProSample. Please refer to the wire colors below when connecting the flow signal
cable to the digital and/or analog device.
NOTE:
The above table is only relevant if item #630147 is used.
NOTE:
An analog and digital device can be connected to the ProSample using the flow signal cable.
NOTE:
The green and common (grey) can be connected to a relay connection. When the relay closes, an event is
triggered and sampling will start.
Wiring and Connecting Analog, Digital I/O,
and SDI-12 Devices
Wire Color
Reference Number
Input Signal
Purpose
Brown
1
Connection of an analog device
White
2
Analog -
Connection of an analog device
Grey
3
Com
Connection of a digital device
Yellow
4
Digital
Connection of a digital device
Green
5
Event
Event
Wiring and Connecting Analog, Digital I/O, and SDI-12 Devices