![YSI MultiDO 5400 Скачать руководство пользователя страница 56](http://html.mh-extra.com/html/ysi/multido-5400/multido-5400_user-manual_915316056.webp)
YSI 5X00
111
YSI 5X00
110
Instrument
Menus → System → Configure E-Aux →E-Aux 1-8→Instrument>Device Address
menu opens
Select the IOEM slave device address whose input will be used as an input signal
for this E-Aux system.
1.
Use
▲
▼
to scroll and highlight IOEM Slave device address.
2.
Press
to select.
3. Press
to save settings and return to Run Screen.
Notes:
-
The Device Address menu displays a generic list of instruments being Master
and Slaves 1-31.
-
The Local Machine user configured name is displayed if the
local machine
is configured.
- See page 86 for local versus remote instrument description.
- The Device Address menu will always display a Master and slaves 1-31
regardless of the actual number of slaves wired to the node network.
Configuring the 5X00
Channel
Menus → System → Configure E-Aux →E-Aux 1-8→Channel>Channel menu opens
Select the IOEM channel whose wired input location will be used as an input signal
for this E-Aux system.
1.
Use
▲
▼
to scroll and highlight IOEM channels.
2.
Press
to select.
3. Press
to save settings and return to Run Screen.
Notes:
-
The Channel menu displays a generic list of instruments being Channel 1-8.
- The Channel menu will always display Channels 1-8 regardless if the
IOEM device selected is an IOEM-4 which has only 4 channels and not 8.
-
IOEM channel modes can only be configured using AquaManager. See Aqua-
Manager online help for directions to configure IOEM channel input mode
(digital, 0-1V, 0-5V, 4mA-20mA, temperature);
Input Type
Menus → System → Configure E-Aux →E-Aux 1-8→Input Type→Digital
→0-1v
→0-5v
→4/20ma
→Temperature
Select the input type for the device(s) wired to IOEM channel(s). E-Aux Sensor Setup
menu selections are based on the E-Aux input type configured at system>E-Aux
menu. Input types are discussed in the following sections.
Configuring the 5X00