Ethernet
Menus → System → Communications → Ethernet
The serial to ethernet device speed and format are self configured at power up (TCP/
IP connections). If the serial to ethernet device is not installed or is not installed
properly, the ethernet menu will display (Not Available).
Notes:
-
5X00 email alarming from ethernet module requires access to an SMTP server
accessible by the local network.
-
Do not reconfigure internal serial to ethernet device settings.
1. Install serial to ethernet device - page 68.
2. The Ethernet icon
displays at the Run Screen when the serial to
ethernet device is properly installed.
3. Enter Ethernet menu
Menus → System → Communications → Ethernet
4.
Use
▲
and
▼
to scroll and highlight submenus.
5.
Press
to select submenu.
6. Configure submenus as needed.
7. Press
to save configuration and return to Run Screen.
Notes:
-
The CC/Email address must be configured when using 5X00 email alarming.
Specific alarm system information will not be sent if the CC/Email address
menu is not configured with a valid email address.
-
If Autopoller is being used for AquaManager email alarming on a TCP/IP
connection, it is strongly recommended that the 5X00 email alarming using
the serial to ethernet device be disabled. When the ethernet device sends
email alarms, all Autopoller SMS communication is suspended. This could
potentially delay any AquaManager alarm emails from being sent.
-
To disable the serial to ethernet device from sending alarms, clear all four
Configuring the 5X00
email addresses from the
Communication → Ethernet
menus. Clear the email
addresses by using the “Spc” (space) key on the numeric keypad. Save changes
at numeric keypad and press
to save configuration and return to Run
Screen.
DHCP
Menus → System → Communications → Ethernet → DHCP
Enable DHCP if you are using the serial to ethernet device only for 5X00 email
alarming and not for AquaManager remote TCP/IP access.
Notes:
-
When DHCP is enabled, the 5X00 IP address is not updated at the IP address
menu.
-
With DHCP enabled, the serial to ethernet device acquires SMTP and DNS
addresses during initialization.
-
AquaManager requires a static IP address; therefore, DHCP must be disabled
when using AquaManager. Consult the network administrator to obtain a
static IP address.
-
Power cycle the 5X00 if DHCP configuration changes from disabled to en-
abled..
IP Address
Menus → System → Communications → Ethernet → IP Address
The IP address can only be configured when DHCP is disabled.
Notes:
-
AquaManager users should program the IP Address either by force serial port
using AquaManager or locally at the 5X00. Poll node using RS232 commu-
nication . Next upload configuration information including: static IP, port,
subnet, and gateway addresses.
Configuring the 5X00