DO Change/pH Change/Temp Change
Enter the percent reduction in feed weight that will occur for “out of range” enabled
feed timer sensors. Results from all out of range sensors are added. To disable a sensor
value from affecting feed duration configure 0.0% as the % change for that sensor.
Calibrate
Menus →Timers→Feed Smart Timer→Feeder→Calibrate
Calibrating the feeder is essential for proper operation of the feed timer system.
Feed timer calibration is required to establish the amount of food dispensed over
a certain period of time. The 5200A will base feed cycle duration times on the last
saved calibration value.
Calibrate Feed Smart Timer
1. Install and wire timer devices - page 46.
OR
1. Wire relay output device(s) to REM - see IOEM/REM Safety and Instal-
lation guide>Chapter 3>Step 4>Wire REM Relay Outputs.
2. Enable Feed Timer.
Menus → Timers→Feed Smart Timer →Feed Timer →Enabled
3. Assign feed timer relay.
Menus → Timers→Feed Smart Timer →Relay
4. Calibrate feeder.
Menus → Timers→Feed Smart Timer →Calibrate
5. At numeric keypad, enter test run time the feeder relay will energize.
6. Press
for feed timer relay to energize for configured run test time.
Countdown of relay “on” time is displayed.
7. When run time expires, the relay de-energizes. Enter the amount of food
dispensed at the Measured Weight screen.
8. Press
to save settings and return to Run Screen.
Notes:
-
Since feeders and food size vary, run several calibrations and enter the aver-
age weight at the last calibration. If feeder or feed type is changed, recalibrate
the feeder.
-
If no test/calibration has been performed or if the results have not been saved,
the 5200A will be unable to perform any feeding(s) because there will be no
reference to determine the rate at which feeding(s) will occur.
-
Stored calibration data is available from the AquaManager program.
-
If feed timer units change, the feeder must be recalibrated.
-
If an incorrect value is entered and saved, another calibration must be per-
formed.
-
Disable the feed timer system when performing the feed timer calibration,
de-energizing any active feed timer relay.
-
After calibration is complete, press
to Run Screen. The feed timer
system resets based on the new calibration configuration.
-
Feed Timer calibration cannot be done using AquaManager.