YSI 5200A
139
YSI 5200A
138
Reset Clean Probe timer at the
menu or the Reset Clean Probes menu.
When prompted “Are you sure?”, select “Yes” to reset the clean probe timer to the
user configured number of days.
Reset To Factory
Menus → System → Service→ Reset To Factory
The Reset to Factory menu resets some, or all, of the 5200A systems to factory
defaults. For example, if the Clean Probe Timer is user configured to 60 days,
performing a factory reset will reset the Clean Probe Timer to 90 days which is the
factory default setting and not the user configured value. See Appendix B Menu
Map
s
- starting on page 226 for factory default settings for all systems.
Notes:
-
“Reset All” and “Reset All Systems” resets time format to 12 hour and date
format to MMDDYYYY.
-
It is strongly recommended that user configuration be maintained in a log
by the user.
-
Instrument, aux, sensor, and suffix names are not reset at any reset to factory
menus. Relay names are reset at “reset relays” and “reset all” menus.
-
Perform sensor calibration after selecting “Reset All” and/or “Reset Sensors”.
Configuring the 5200A
-
Control icons and control menu messages are displayed for 30 seconds after
pressing
from reset to factory menus for any system that was active
(energized relay) prior to the factory reset.
Reset All
Menus → System → Service→ Reset To Factory→ Reset All
Resets all systems, sensors, and timers to factory defaults.
1. Use
▲
and
▼
to scroll and highlight submenus.
2. Press
to select.
3. When prompted, “Are you sure?”, select one of the following to confirm,
4. Press
to save configuration and return to Run Screen.
Notes:
-
Display contrast, backlight timer, and buzzer can only be reset to factory
defaults through this menu.
-
Sensors must be recalibrated after performing a factory Reset All and/or
Reset Sensor.
Reset System
Menus → System → Service→ Reset To Factory→ Reset System
Specific systems or all systems listed in the Reset System menu can be reset to fac-
tory defaults.
1. Use
▲
and
▼
to scroll and highlight submenus.
2. Press
to select.
Configuring the 5200A