YSI Incorporated
Model 5000/5100 Service Manual
11
98.0
*
% 8.15
mg/L
Calibrate
0.0
ppt
740
mmHg
24.38°C 11:33
AM
AUTO CAL
DO CAL
SETUP
DIAGNOSTICS
7.
Model 5100:
Verify that the barometer reading is correct. Calibrate the barometer, if
necessary, as shown under
Calibration, Barometer
.
Model 5000:
Enter the current barometric pressure as shown under
Calibration, Barometer
.
8. Verify that the salinity is set to the correct value (0.0 ppt for air calibration). See
Calibration,
Salinity
.
Make sure that the display readings are stable, then press the [AUTO CAL] soft-key to calibrate
Dissolved Oxygen. The message “
D.O. CALIBRATION SAVED
” will be displayed for a few
seconds.
98.0
*
% 8.15
mg/L
Calibrate
0.0
ppt
740
mmHg
24.38°C 11:33
AM
D. O. CALIBRATION SAVED
AUTO CAL
DO CAL
SETUP
DIAGNOSTICS
Press [MODE] to return to the Main mode. The instrument is now calibrated and ready to
measure dissolved oxygen and temperature. See
Operation, Main Mode, Making Measurements
.
DO CAL (MANUAL)
1. Prepare the probe according to the probe instructions.
2. Connect the probe to the meter.
3. Place the probe in a known oxygen environment, such as a BOD bottle containing about 1
″
of water to provide a 100% relative humidity environment or a Winkler-titrated sample.
4. Press [ ] to turn the instrument on.
5. Allow the probe to polarize and the temperature to stabilize for at least 15 minutes. If
calibration is performed prematurely the values will drift and may be out of specification.
6. Press the [CALIBRATE] soft-key to change to Calibration mode. The following screen will
be displayed.