32
M Soften
Resume
Line correlation is the processing of scan lines of M images.
Effect
Suppress noise to make image details clearer.
M line
Resume
Adjust the position of the M sampling line.
4.5
Parameters adjustment for Color
mode
Color Doppler is used to observe the color blood flow and provide the flow direction and
velocity information of the blood flow. In general, red indicates the blood flow to the probe,
blue indicates the blood flow away from the probe, the brighter the color, the faster the
blood flow, the darker the color, and the slower the blood flow rate.
WARNING 1.
Color Doppler flow can only be used as a reference for
doctors and can not be diagnosed directly. It is generally
compared with other machines or diagnosed by non-
ultrasonic means.
2. This device cannot perform long-term examination of the
fetus in Doppler mode.
4.5.1
Color mode image parameter
During C-mode image scanning, the image parameter area on the left side of the screen
provides real-time parameter information, which allows you to adjust the following
parameters:
Adjustment method
Parameter term
Control panel adjustment
Gain, depth
SoftKey adjustment
Velocity, deflection, Frequency, Color first, baseline,
atlas, Wall filter, Sound Power, afterglow
The Color mode shares the probe sound power of the B mode ,When the acoustic power
is adjusted, the sound power of the two imaging modes changes synchronously.
Because the Color mode is superimposed by two-dimensional (B) mode and Doppler mode,
the two kinds of images change synchronously when magnifying or adjusting the depth.