•The thermometer contains sensitive elements, which should be treated with caution. The aging of
the elements will affect the performance of the thermometer. At the same time, the thermometer
should not be used in the environment with electromagnetic interference.
•The infrared thermometer is composed of high quality sensitive parts, which can not be repaired
by users themselves.
•Please place the thermometer in a place where children, pets, etc. can not be touched, so as to
prevent falling, biting and affecting the product performance. Children need to use it under the
supervision of adults.
•Warning: this device is not allowed to be modified
NOTE: Please follow local laws to dispose of waste scrap.
•The results of ear temperature measurement by infrared thermometer were compared with those
of oral mercury thermometer in control group.
♦Preparation before use:
Before the controlled trial, all subjects had no eating, drinking or strenuous activity for half an
hour, and the environmental conditions were referred to the instruction.
♦Inspection method
:
During the measurement, the mercury end of the oral thermometer, which was thrown below 35
℃
,
was placed at an Angle in the hot bag under the subject's tongue, and the body temperature was
read and recorded after 5 minutes with the lips closed tightly.Let's say the mercury thermometer is
37.5
℃
.
At the same time as the oral temperature is measured, the ear temperature is measured.The user
reads three consecutive temperatures in the same ear for no more than one minute at a time. The
average value is taken, if the result is a reading of 37.2
℃
, compared with the value of the mercury
thermometer.The difference between the temperature of 37.5
℃
and the mercury thermometer is
within the error range of ±0.3
℃
, indicating that the infrared thermometer is normal.
If the reading is 36.5
℃
, with a difference of 1
℃
, the infrared thermometer needs to be calibrated.
measures
You can contact the manufacturer for calibration inspection within the warranty period.
The calibration cycle of this equipment is 2 years, and the calibration unit is: the manufacturer.
•The verification of the basic safety and performance of the thermometer shall comply with the
inspection procedures for the whole infrared thermometer.
No.
Item
1
Name
Infrared Thermometer