<8. Detailed Data Setting>
8-5
8.2.4
Default Values
When the analyzer is delivered, or if data are initialized, output hold is the default as shown in
Table 8.3.
Table 8.3 Output Hold Default Values
Status
Output hold setting
Preset value
During warm-up
4 mA
4 mA
Under maintenance
Holds output at value just before maintenance started
4 mA
Under calibration or blow back Holds output at value just before starting calibration or
blow back
4 mA
On Error occurrence
Holds output at a preset value
3.4 mA
IM 11M13A01-02E