<7. Startup>
7-15
IM 11M13A01-04E
7.10 Calibration
The analyzer is calibrated in such a way that the actual zero and span gases are measured and those
measured values are used to agree with the oxygen concentrations in the respective gases.
There are three types of calibration procedures available:
(1) Manual calibration conducting zero and span calibrations, or either of these calibrations in turn.
(2) Semi-automatic calibration which uses the infrared switches or a contact input signal and
conducts calibration operations based on a preset calibration time and stable time.
(3) Automatic calibration conducted at preset intervals.
Manual calibration needs the ZA8F Flow Setting Unit to allow manual supply of the calibration gases.
Semi-automatic and automatic calibrations need Automatic Calibration Unit to allow automatic supply
of the calibration gases. The following sections set forth the manual calibration procedures. For details
on semi-automatic and automatic calibrations, consult Chapter 9, “Calibration” later in this manual.
7.10.1 Calibration Setup
Set the following three items before carrying out a calibration. Parameter codes for these set items are
listed in Table 7.13.
(1) Mode setting
There are three calibration modes: manual, semi-automatic, and automatic.
Select the desired mode. This section uses manual mode for calibration.
(2) Oxygen concentration in zero gas
Enter the zero gas oxygen concentration for calibration.
(3) Oxygen concentration in span gas
Enter the span gas oxygen concentration for calibration. If instrument air is used, enter 21 vol%
O
2
. When using the ZO21S Standard Gas Unit (for use of the atmospheric air as a span gas),
use a hand-held oxygen analyzer to measure the actual oxygen concentration, and then enter it.
CAUTION
If instrument air is used for the span gas, dehumidify the air to a dew point of -20°C and remove any
oil mist or dust.
Incomplete dehumidifying or unclean air will have an adverse effect on the measurement accuracy.
Table 7.13
Calibration Parameter Codes
Set item
Parameter code
Set value
Calibration mode
B03
0 Manual calibration
1 Semi-automatic, Manual calibration
2 Automatic, Semi-automatic, Manual calibration
Zero gas oxygen concentration
B01
Enter oxygen concentration.
Span gas oxygen concentration
B02
Enter oxygen concentration.
Содержание ZR202S
Страница 12: ...xi IM 11M13A01 04E Introduction ...
Страница 18: ...Blank Page ...
Страница 61: ...Blank Page ...
Страница 129: ...Blank Page ...
Страница 141: ...Blank Page ...
Страница 153: ......
Страница 155: ......
Страница 157: ......
Страница 159: ...Blank Page ...
Страница 160: ......
Страница 161: ......
Страница 162: ......
Страница 163: ......