<3. Installation>
3-3
IM 11M13A01-04E
Note 7: Special fastener
Hexagon socket head cap screws below are special fastener according to EN60079-0: 2012+A11:
2013 (see the figure below). Material property classes of them are A2-50 or better.
Special fastener
Figure 3.1
Special fastener
3.1.3 FM Explosion-proof Type
ZR202S–B Analyzer for use in hazardous area:
Note 1: Applicable Standard: FM3600 1998, FM3615 1989, FM3810 2005, ANSI/NEMA 250 1991
Type of protection:
Explosion-proof for Class I, Division 1, Groups B, C and D
Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G
Enclosure Rating:
NEMA 4X
Temperature Class:
T2
Note 2: Wiring
• All wiring shall comply with National Electrical Code ANSI/NEPA 70 and Local Electrical Code.
• In hazardous area, wiring shall be in conduits as shown in the figure.
WARNING:
SEAL ALL CONDUITS WITHIN 18 INCHES OF THE ENCLOSURE.
Hazardous Area Division 1
18 inches (475mm) MAX.
18 inches (475mm) MAX.
Non-hazardous
Area
Non-hazardous
Area
Conduit
Sealing Fitting
F3-1E.ai
Figure 3.2 Wiring of ZR202S-B