<3. Installation>
3-1
IM 11M13A01-04E
3. Installation
This chapter describes installation of the following equipment:
3.1 Model ZR202S Integrated type Explosion-proof Zirconia Oxygen Analyzer
3.2 Model ZA8F Flow Setting Unit
3.3 Case Assembly (E7044KF) for Calibration gas Cylinder
3.1 Installation of ZR202S Zirconia Oxygen Analyzer
3.1.1 Installation Location
The following should be taken into consideration when installing the analyzer:
(1) Easy and safe access to the analyzer for checking and maintenance work.
(2) Ambient temperature of not more than 55°C, and the terminal box should not be affected by
radiant heat.
(3) A clean environment without any corrosive gases.
(4) No vibration.
(5) The sample gas satisfies the specifications described in Chapter 2.
(6) No sample gas pressure fluctuations.
CAUTION
• The ambient temperature of the ZR202S Integrated type Explosion-proof Zirconia Oxygen
Analyzer should be between - 20°C and 55°C.