<12. Troubleshooting>
12-4
IM 11M13A01-04E
12.2 Displays and Remedies When Alarms are
Generated
12.2.1 Alarm types
When an alarm is generated, the alarm indication blinks in the display to notify of the alarm (Figure 12.4).
Alarms include those shown in Table 12.2.
Displayed alternately
AL-06
21.0%
F12-4E.ai
Figure 12.4
Table 12.2
Types of Alarms and Reasons for Occurrence
Alarm No.
Type of alarm
Reason for occurrence
AL-01
Oxygen concentration alarm
Occurs when a measured value exceed or falls below
the set alarm value (refer to Section 8.3, “Alarm Setting”).
AL-06
Zero calibration coefficient
(correction ratio) alarm
Generated when the zero correction ratio is out of the range of
100±30% in automatic and semi-automatic calibration (refer to
Subsection 9.1.3, Compensation).
AL-07
Span calibration coefficient
(correction ratio) alarm
Generated when the span correction ratio is out of the range
of 0±18% in automatic and semi-automatic calibration (refer to
Subsection 9.1.3, “Compensation”).
AL-08
EMF stabilization time-up alarm
Generated when the cell (sensor) voltage is not stabilized
even after the calibration time is up in automatic and semi-
automatic calibration.
AL-10
Cold junction temperature alarm
Occurs when an equipment internal temperature
exceeds 85°C.
AL-11
Thermocouple voltage alarm
Generated when thermocouple voltage exceeds 42.1 mV
(about 1020°C ) or falls below -5 mV (about -170°C).
AL-13
Battery low alarm
Internal battery needs replacement.
If an alarm is generated, actions such as turning off the heater power are not carried out. The alarm is
released when the cause for the alarm is eliminated.
However, AL-10 and/or AL-11 may be generated at the same time as Err-02 (heater temperature
failure).
In such a case, the measure taken for this error has priority.
If the converter power is turned off after an alarm is generated and restarted before the cause of the
alarm has been eliminated, the alarm will be generated again.
However, AL- 6, 7, and 8 (alarms related to calibration) are not generated unless calibration is
executed.
Содержание ZR202S
Страница 12: ...xi IM 11M13A01 04E Introduction ...
Страница 18: ...Blank Page ...
Страница 61: ...Blank Page ...
Страница 129: ...Blank Page ...
Страница 141: ...Blank Page ...
Страница 153: ......
Страница 155: ......
Страница 157: ......
Страница 159: ...Blank Page ...
Страница 160: ......
Страница 161: ......
Страница 162: ......
Страница 163: ......