<10. Other Functions>
10-10
IM 11M13A01-04E
10.4 Reset
Resetting enables the equipment to restart. If the equipment is reset, the power is turned off and
then back on. In practical use, the power remains on, and the equipment is restarted under program
control. Resetting will be possible in the following conditions:
(1) Err-01
if the cell voltage is defective
(2) Err-02
if a temperature alarm occurs
(3) Err-03
if the A/D converter is defective
(4) Err-04
if an EEPROM write error occurs
For details on error occurrence, consult Chapter 12, “Troubleshooting”, later in this manual.
If any of the above problems occurs, the equipment turns off the power to the detector heater. To
cancel the error, reset the equipment following the steps below, or turn the power off and then back
on.
Note: Make sure that before resetting or restarting the power that there is no problem with the equipment.
CAUTION
• If a problem arises again after the resetting, turn the power off and troubleshoot the problem by
consulting the Troubleshooting chapter later in this manual. When there is no error, the Basic
panel display will appear.