<3. PIPING>
3-3
IM 11M03A02-00E
3.1.3
Piping for the Reference Gas Inlet
Reference gas piping is required between the air source (fresh air) and the flow setting unit and
between the flow setting unit and the detector.
Insert the air set next to the flow setting unit in the piping between the air source and the flow
setting unit.
Use a ø6 × ø4 mm (or nominal size 1/4 inch) stainless steel pipe between the flow setting unit
and the detector.