6.2 Adjustment
• For details of operation and adjusting procedures of VJ77 Parameters Setting Tool,
refer to the instruction manual “Model VJ77 PC-based Parameters Setting Tool” (IM
77J01J77-01E).
• For details of operation and adjusting procedures of JHT200 Handy Terminal, refer
to the instruction manual ‘‘JHT200 Handy Terminal’’ (IM 77J50H01-01EN).
• Do not turn off the power of the instrument during adjustment.
Use a PC (VJ77) or the JHT200 Handy Terminal for adjustment.
The procedure is shown below using the JHT200 Handy Terminal as an example.
Adjust SISD using JHT200 Handy Terminal.
(a) Connect the equipment as illustrated in Figure 6-1.
(b) Set the Parameter Write Protect (W.P.) of setting jumper to OFF. (Refer to ‘‘5.2 Setting
Jumper’’.)
(c) Turn on the power, and allow the equipment to warm up for about 5 minutes under this
condition.
(d) Connect JHT200 Handy Terminal.
F0601.ai
<Connection>
Adapter for modular jack
(E9786WH)
BRAIN
connector
JHT200
Handy Terminal
BRAIN
connector
Dedicated adapter
for VJ77 (L4506HA)
PC
MicroUSB (USB2.0) cable
JUXTA communication cable
JUXTA communication cable
Modular-jack
conversion adapter
Figure
6-1
Connection