2-26
IM WT5000-02EN
2.7 Setting the Line Filter and Frequency Filter
Second Frequency Measurement Settings (Freq2 Measurement)
3.
Tap
Freq Filter/Rectifier/Level
. A frequency filter setup screen appears.
4.
Tap
Freq2 Measurement
. A Freq2 Measurement (second frequency measurement) screen
appears.
Input (Advanced/Options) tab
Set the second frequency measurement.
By tapping Cutoff, Voltage Level (Freq2), or
Current Level (Freq2), you can set the value
collectively for all elements.
By tapping Freq Filter (Freq2), you can
turn on or off the featur on all elements
collectively.
Set the cutoff frequency of the high-pass filter (HPF)
(0.1 Hz, 1 Hz, 10 Hz, 0.1 kHz to 100.0 kHz (0.1 kHz steps)).
Turns the high-pass filter
(HPF) on or off
Set the voltage
cross level (0.0% to
100.0%(0.1% steps)).
You can use the voltage and current cross
level settings when the high-pass filter (HPF)
is set to OFF.
Set the current
cross level (0.0% to
100.0%(0.1% steps)).
Procedure Using the Menu Icons
You can also use the menu icons shown on the right side of the screen to set the line filter and
frequency filter.
1.
Tap the
Filter
menu icon
. A Filter menu appears in the sub menu area on the right side of
the screen.
By tapping the displayed items, you can specify the same settings as when using the keys explained
earlier.
Note
For details on the Filter menu, see page x.