16-11
IM WT5000-01EN
Storage Operations in Each Storage Mode
Manual (Manual)
Time
REC
key
State
REC
REC
Solid
CMPL
Storage count
PAUSE
Solid
END
key
Real-Time Control (Real Time)
Time
Scheduled
start time
REC
Blinking
REC
key
State
READY
REC
REC
Solid
CMPL
Scheduled
stop time or
storage count
PAUSE
Solid
END
key
Integration Synchronization (Integ Sync)
When the storage count is reached before integration stops
Time
Integration
start
REC
Blinking
REC
key
State
READY
REC
REC
Solid
CMPL
Storage
count
PAUSE
Solid
END
key
When integration stops before the storage count is reached (illustration of operations after integration starts)
Time
REC
Blinking
State
READY
REC
REC
Solid
PAUSE
Integration
reset
PAUSE
Blinking
END
key
Integration
start
Integration
stop
• If you reset integration after it has been stopped, the store state changes to complete (CMPL). Writing to the
stored data file (.wts) and header file (.hds) finishes, and the files are closed.
• If you restart integration after it has been stopped, the store state changes to recording (REC). The stored
data continues to be written to the file from before storage was paused.
Event Synchronization (Event)
Time
Event occurrence
REC
Blinking
REC
key
State
READY
REC
REC
Solid
CMPL
Storage count
PAUSE
Solid
END
key
16 Data Storage