![YOKOGAWA WT5000 Скачать руководство пользователя страница 136](http://html1.mh-extra.com/html/yokogawa/wt5000/wt5000_features-manual_4140530136.webp)
14-13
IM WT5000-01EN
Starting, Stopping, and Resetting Integration
You can start, stop, and reset integration using the touch panel buttons, front panel keys, or communication
commands. The relationships between integration and starting, stopping, and resetting are indicated below.
Integrated
value
Integration
time
Timeout
Timer value
Start
Stop
Start
Stop
Reset
Start
Reset
Starting Integration (Start, START)
• Integration starts under the different conditions indicated below depending on the integration mode.
• Manual integration mode, standard integration mode, and continuous integration mode
Integration starts immediately.
• Real-time normal integration mode and real-time continuous integration mode
This instrument enters into an integration-ready state. Integration starts when the scheduled start time is
reached.
• When integration starts, the START key illuminates, and the integration status shows .
• When this instrument enters an integration-ready state, the START key blinks, and the integration status
shows .
Stopping Integration (Stop, STOP)
• You can force integration to stop. The integration time and integrated value are held.
• When you force integration to stop, the STOP key blinks, and the integration status shows . Tapping Start
or pressing START resumes the integration. However, after the automatic stopping of integration, which is
discussed later, you cannot tap Start or press START to resume integration.
Automatic Stopping of Integration
• Integration stops automatically under the different conditions indicated below depending on the integration
mode. The integration time and integrated value are held. After integration stops automatically, you cannot
resume it by tapping Start or pressing START. To start integration, reset it first.
• When the integration mode is manual integration mode, standard integration mode, or real-time normal
integration mode
After the specified timer time elapses, integration stops automatically. The STOP key illuminates, and the
integration status shows .
• When the integration mode is real-time normal integration mode or real-time continuous integration mode
Integration stops automatically when the scheduled integration end time is reached. The STOP key
illuminates, and the integration status shows .
14 Integrated Power (Watt Hours/Ampere Hours)