Description In the parameters, list the external current sensor
ranges that you want to enable. To enable all the
ranges, specify the parameter “ALL.”
[:INPut]:CURRent:EXTSensor:POJump
Function
Sets or queries the jump destination range that is
used when a current peak over-range occurs.
Syntax
[:INPut]:CURRent:EXTSensor:POJump
{OFF|<
Voltage
>}
[:INPut]:CURRent:EXTSensor:POJump?
OFF
= No jump destination current range
<Voltage> =
See (
:INPut:CURRent:
RANGe
).
Example
:INPUT:CURRENT:EXTSENSOR:POJUMP 2V
:INPUT:CURRENT:EXTSENSOR:POJUMP?
-> :INPUT:CURRENT:EXTSENSOR:POJUMP
2.00E+00
[:INPut]:CURRent:SRATio?
Function
Queries the external
current sensor conversion
ratios of all elements.
Syntax
[:INPut]:CURRent:SRATio?
Description This command is only valid on models with the
external current sensor input (/EX1 or /EX2)
option.
[:INPut]:CURRent:SRATio[:ALL]
Function
Collectively sets the external current sensor
conversion ratios of all elements.
Syntax
[:INPut]:CURRent:SRATio[:ALL]
{<NRf>}
<NRf>
=
0.001
to
9999.
Example
:INPUT:CURRENT:SRATIO:ALL 10
[:INPut]:CURRent:SRATio:ELEMent<x>
Function
Sets or queries the external current sensor
conversion ratio of the specified element.
Syntax
[:INPut]:CURRent:SRATio:ELEMent<x>
{<NRf>}
[:INPut]:CURRent:SRATio:ELEMent<x>?
<x>
=
1
to
3
(element)
<NRf>
=
0.001
to
9999.
Example
:INPUT:CURRENT:SRATIO:ELEMENT1 10
:INPUT:CURRENT:SRATIO:ELEMENT1?
-> :INPUT:CURRENT:SRATIO:ELEMENT1
10.00
[:INPut]:RCONfig
Function
Sets or queries the on/off state of the
range
configuration (valid range selection) feature.
Syntax
[:INPut]:RCONfig {<Boolean>}
[:INPut]:RCONfig?
Example
:INPUT:RCONFIG OFF
:INPUT:RCONFIG? -> :INPUT:RCONFIG 0
Description The following commands are enabled only when
this command is set to ON. Measurement range
can be skipped.
:INPut:VOLTage:CONFig
:INPut:VOLTage:POJump
:INPut:CURRent:CONFig
:INPut:CURRent:POJump
:INPut:CURRent:EXTSensor:CONFig
:INPut:CURRent:EXTSensor:POJump
[:INPut]:SCALing?
Function
Queries all
scaling settings.
Syntax
[:INPut]:SCALing?
[:INPut]:SCALing[:STATe]
Function
Sets or queries the scaling on/off state.
Syntax
[:INPut]:SCALing[:STATe] {<Boolean>}
[:INPut]:SCALing:STATe?
Example
:INPUT:SCALING:STATE OFF
:INPUT:SCALING:STATE?
-> :INPUT:SCALING:STATE 0
[:INPut]:SCALing:{VT|CT|SFACtor}?
Function
Queries the VT ratios,
CT ratios, or power
coefficients of all elements.
Syntax
[:INPut]:SCALing:{VT|CT|SFACtor}?
[:INPut]:SCALing:{VT|CT|SFACtor}
[:ALL]
Function
Collectively sets the VT ratio, CT ratio, or power
coefficient of all elements.
Syntax
[:INPut]:SCALing:{VT|CT|SFACtor}
[:ALL] {<NRf>}
<NRf>
=
0.001
to
9999.
Example
:INPUT:SCALING:VT:ALL 1
[:INPut]:SCALing:{VT|CT|SFACtor}:
ELEMent<x>
Function
Sets or queries the VT ratio, CT ratio, or power
coefficient of the specified element.
Syntax
[:INPut]:SCALing:{VT|CT|SFACtor}
:ELEMent<x> {<NRf>}
[:INPut]:SCALing:{VT|CT|SFACtor}
:ELEMent<x>?
<x>
=
1
to
3
(element)
<NRf>
=
0.001
to
9999.
Example
:INPUT:SCALING:VT:ELEMENT1 1
:INPUT:SCALING:VT:ELEMENT1?
-> :INPUT:SCALING:VT:ELEMENT1 1.000
[:INPut]:SYNChronize
Function
Sets or queries the
synchronization source.
Syntax
[:INPut]:SYNChronize
{VOLTage|CURRent|OFF}
[:INPut]:SYNChronize?
Example
:INPUT:SYNCHRONIZE VOLTAGE
:INPUT:SYNCHRONIZE?
-> :INPUT:SYNCHRONIZE VOLTAGE