6-13
IM WT310E-17EN
Commands
1
2
3
4
5
6
7
8
9
App
Index
[:INPut]:VOLTage:POJump
Function
Sets or queries the jump destination range that is
used when a voltage peak over-range occurs.
Syntax
[:INPut]:VOLTage:POJump
{OFF|<
Voltage
>}
[:INPut]:VOLTage:POJump?
OFF
= No jump destination voltage range
<Voltage> =
See (
:INPut:VOLTage:RANGe
).
Example
:INPUT:VOLTAGE:POJUMP 600V
:INPUT:VOLTAGE:POJUMP?
-> :INPUT:VOLTAGE:POJUMP 600.0E+00
[:INPut]:CURRent?
Function
Queries all electric current measurement settings.
Syntax
[:INPut]:CURRent?
[:INPut]:CURRent:RANGe
Function
Sets or queries the
current range.
Syntax
[:INPut]:CURRent:RANGe
{<
Current
>|(EXTernal,<
Voltage
>)}
[:INPut]:CURRent:RANGe?
• For direct current input
• When the crest factor is set to 3
<Current> = 5, 10, 20, 50, 100, 200, 500(mA),
1, 2, 5, 10, 20(A) (WT310E)
<Current> = 1, 2, 5, 10, 20, 40(A) (WT310EH)
<Current> = 0.5, 1, 2, 5, 10, 20(A) (WT332E,
WT333E)
• When the crest factor is set to 6 or 6A
<Current> = 2.5, 5, 10, 25, 50, 100, 250(mA),
0.5, 1, 2.5, 5, 10(A) (WT310E)
<Current> = 0.5, 1, 2.5, 5, 10, 20(A) (WT310EH)
<Current> = 0.25, 0.5, 1, 2.5, 5, 10(A) (WT332E,
WT333E)
• For external current sensor input
• When the crest factor is set to 3
<Voltage> = 2.5, 5, 10(V) (/EX1)
<Voltage> = 50, 100, 200, 500(mV), 1, 2(V)
(/EX2)
• When the crest factor is set to 6 or 6A
<Voltage> = 1.25, 2.5, 5(V) (/EX1)
<Voltage> = 25, 50, 100, 250(mV), 0.5, 1(V)
(/EX2)
Example
:INPUT:CURRENT:RANGE 20A
:INPUT:CURRENT:RANGE?
-> :INPUT:CURRENT:RANGE 20.0E+00
:INPUT:CURRENT:RANGE EXTERNAL,10V
:INPUT:CURRENT:RANGE?
-> :INPUT:CURRENT:RANGE
EXTERNAL,10.0E+00
Description EXTernal and <Voltage> can only be selected
on models with the external current sensor input
(/EX1 or /EX2) option.
[:INPut]:CURRent:AUTO
Function
Sets or queries the current
auto range on/off state.
Syntax
[:INPut]:CURRent:AUTO {<Boolean>}
[:INPut]:CURRent:AUTO?
Example
:INPUT:CURRENT:AUTO ON
:INPUT:CURRENT:AUTO?
-> :INPUT:CURRENT:AUTO 1
[:INPut]:CURRent:CONFig
Function
Sets or queries the valid current range.
Syntax
[:INPut]:CURRent:CONFig {ALL|<
Current
>
[,<
Current
>][,<
Current
>]...}
[:INPut]:CURRent:CONFig?
ALL
= All ranges are valid.
<Current> =
See (
:INPut:CURRent:RANGe
).
Example
:INPUT:CURRENT:CONFIG ALL
:INPUT:CURRENT:CONFIG?
-> :INPUT:CURRENT:CONFIG ALL
:INPUT:CURRENT:CONFIG 20,5,1
:INPUT:CURRENT:CONFIG?
-> :INPUT:CURRENT:CONFIG
20.0E+00,5.0E+00,1.0E+00
Description In the parameters, list the current ranges that you
want to enable. To enable all the ranges, specify
the parameter “ALL.”
[:INPut]:CURRent:POJump
Function
Sets or queries the jump destination range that is
used when a current peak over-range occurs.
Syntax
[:INPut]:CURRent:POJump
{OFF|<
Current
>}
[:INPut]:CURRent:POJump?
OFF
= No jump destination current range
<Current> =
See (
:INPut:CURRent:RANGe
).
Example
:INPUT:CURRENT:POJUMP 20A
:INPUT:CURRENT:POJUMP?
-> :INPUT:CURRENT:POJUMP 20.0E+00
[:INPut]:CURRent:EXTSensor:CONFig
Function
Sets or queries the valid external
current sensor
range.
Syntax
[:INPut]:CURRent:EXTSensor:CONFig
{ALL|<
Voltage
>[,<
Voltage
>]
[,<
Voltage
>]...}
[:INPut]:CURRent:EXTSensor:CONFig?
ALL
= All ranges are valid.
<Voltage> =
See (
:INPut:CURRent:RANGe
).
Example
:INPUT:CURRENT:EXTSENSOR:CONFIG ALL
:INPUT:CURRENT:EXTSENSOR:CONFIG? ->
:INPUT:CURRENT:EXTSENSOR:CONFIG ALL
:INPUT:CURRENT:EXTSENSOR:CONFIG
2,0.5,0.1
:INPUT:CURRENT:EXTSENSOR:CONFIG?
-> :INPUT:CURRENT:EXTSENSOR:CONFIG
2.00E+00,500.0E-03,100.0E-03
6.7 INPut Group