IM WT310E-01EN
5.3 Displaying Integrated Values
Procedure
Displayed Function
Press the
FUNCTION
key for display A or C to select TIME (elapsed integration time), Wh/Wh±
(integrated power), Ah/Ah± (current integration), or MATH (average active power during integration).
Each time you press FUNCTION, the displayed function changes in the order shown below.
TIME
Display
A
C
Wh
Wh±(+)
Ah±(-)
Ah±(+)
Ah
Wh±(-)
MATH
• Vpk, Apk, Wpk, Wh±, and Ah± all illuminate twice. The first time is for the positive measurements and the
second time is for the negative measurements.
• The MATH indicator is on the left side of the 7-segment LED display.
• Press SHIFT before pressing FUNCTION to change the displayed function in reverse order.
Input Element (WT332E/WT333E only)
Press the
ELEMENT
key for display C to select which element to display.
The procedure for selecting an input element is the same as the procedure in section 4.1.
The WT310E and WT310EH only have one input element, so there is no input element selection.
Starting and Stopping Integration
1.
Press
START
.
The START key illuminates, and integration starts. The integrated value (or the average active power during
integration) is shown on display C, and the elapsed integration time is shown on display A.
2.
Press
HOLD
.
The HOLD key illuminates, and the displayed value is held.
3.
Press
HOLD
when the measured values are being held.
The HOLD key light turns off, the held state is released, and the display of the integration results is
updated.
4.
Press
SHIFT+HOLD
(SINGLE) when the measured values are being held.
This instrument will perform a single measurement, update the display, and hold the new values.
5.
Press
STOP
.
Integration stops. The START key light turns off, and the STOP key illuminates. The displayed integration
value is held.