12-28
IM WT3001E-51EN
• Completing the Measurement of dmax Caused by Manual Switching and
Displaying the Judgement
24.
Check that the measurement of all observation periods (24) is complete, and that
the dmax data of each observation period is displayed.
25.
Press the
Judge
soft key. The dmax data of all observation periods is confirmed,
and the measurement of dmax caused by manual switching is complete. The
flicker measurement status indicates Complete, and the judgement is displayed.
• Re-judging the Measurement of dmax Caused by Manual Switching
26.
When the flicker measurement status is indicating Complete, change the
judgement conditions according to the procedure given in “Setting the Judgement
Conditions” in section 12.4. When you change the judgement conditions, the
measured dmax data is re-judged, and the judgement is updated.
• Switching the Element for Displaying the Measurement/Judgement Data
There are two ways to switch the element for displaying the measurement and
judgement data as described below.
• Changing Using the Function Select Keys
27.
Press
ELEMENT
to set the element to be displayed.
The element to be displayed switches in order as in 1, 2, ..., 1.
• Selecting from the Menu
27.
Press
ITEM
to display the Flicker Items menu.
28.
Press the
cursor keys
to set the element to be displayed.
12.6 Executing the Measurement of dmax Caused by Manual Switching