12-11
IM WT3001E-51EN
Voltage Fluctuation and Flicker Measurement (Option)
1
2
3
4
5
6
7
8
9
10
11
12
13
Index
• Setting the Measurement Count
25.
Press the
cursor keys
to select Count.
26.
Press
SET
to display the entry box.
27.
Press the
cursor keys
to set the measurement count.
28.
Press
SET
or
ESC
to close the box.
• Setting the Steady-State Range (dmin: Allowable Range of Relative
Voltage Change to Be Considered Steady-State)
29.
Press the
cursor keys
to select dmin.
30.
Press
SET
. An entry box appears.
31.
Press the
cursor keys
to set the steady-state range.
32.
Press
SET
or
ESC
to close the box.
Explanation
For an explanation of the terminology, see section 12.1.
• Measurement Method
Select the voltage fluctuation and flicker measurement method from the two methods below.
• Flicker (normal flicker measurement)
Judges the voltage fluctuation and flicker from dc, dmax, Tmax, and Pst.
• dmax (measurement of dmax caused by manual switching)
You manually turn the EUT switch ON. This instrument measures the voltage
fluctuation caused by the inrush current that flows when the power is turned ON,
and judges the voltage fluctuation and flicker from the dmax average.
• Edition Number of IEC 61000-4-15
You can select the edition number of IEC 61000-4-15.
• Edition 1.1
• Edition 2.0
IEC 61000-4-15 specifies requirements for measurement instruments.
12.3 Setting the Measurement Conditions