6-68
IM WT3001E-17EN
[:INPut]:CURRent:RANGe:ELEMent<x>
Function
Sets the current range of the element or queries
the current setting.
Syntax
[:INPut]:CURRent:RANGe:ELEMent<x>
{<Current>|(EXTernal,<Voltage>)}
[:INPut]:CURRent:RANGe:ELEMent<x>?
<x> = 1 to 4 (element)
For the 30 A input element
• When the crest factor is set to 3
<Current> = 500 (mA), 1, 2, 5, 10, 20, 30 (A) (for
direct current input)
<Voltage> = 50, 100, 200, 500 (mV), 1, 2, 5, 10
(V) (for external current sensor input)
• When the crest factor is set to 6
<Current> = 250, 500 (mA), 1, 2.5, 5, 10, 15 (A)
(for direct current input)
<Voltage> = 25, 50, 100, 250, 500 (mV), 1, 2.5,
5 (V) (for external current sensor input)
For the 2 A input element
• When crest factor is set to 3
<Current> = 5, 10, 20, 50, 100, 200, 500 (mA),
1, 2, (A) (with direct current input)
<Voltage> = 50, 100, 200, 500 (mV), 1, 2, 5, 10
(V) (with external current sensor input)
• When crest factor is set to 6
<Current> = 2.5, 5, 10, 25, 50, 100, 250, 500
(mA), 1 (A) (with direct current input)
<Voltage> = 25, 50, 100, 250, 500 (mV), 1, 2.5,
5 (V) (with external current sensor input)
Example
:INPUT:CURRENT:RANGE:ELEMENT1 30A
:INPUT:CURRENT:RANGE:ELEMENT1? ->
:INPUT:CURRENT:RANGE:
ELEMENT1 30.0E+00
:INPUT:CURRENT:RANGE:
ELEMENT1 EXTERNAL,10V
:INPUT:CURRENT:RANGE:ELEMENT1? ->
:INPUT:CURRENT:RANGE:
ELEMENT1 EXTERNAL,10.00E+00
[:INPut]:CURRent:RANGe:{SIGMA|SIGMB}
Function
Collectively sets the current range of all elements
belonging to wiring unit {ΣA|ΣB}.
Syntax
[:INPut]:CURRent:RANGe:{SIGMA|SIGMB}
{<Current>|(EXTernal,<Voltage>)}
For the 30 A input element
• When the crest factor is set to 3
<Current> = 500 (mA), 1, 2, 5, 10, 20, 30 (A) (for
direct current input)
<Voltage> = 50, 100, 200, 500 (mV), 1, 2, 5, 10
(V) (for external current sensor input)
• When the crest factor is set to 6
<Current> = 250, 500 (mA), 1, 2.5, 5, 10, 15 (A)
(for direct current input)
<Voltage> = 25, 50, 100, 250, 500 (mV), 1, 2.5,
5 (V) (for external current sensor input)
For the 2 A input element
• When crest factor is set to 3
<Current> = 5, 10, 20, 50, 100, 200, 500 (mA),
1, 2, (A) (with direct current input)
<Voltage> = 50, 100, 200, 500 (mV), 1, 2, 5, 10
(V) (with external current sensor input)
• When crest factor is set to 6
<Current> = 2.5, 5, 10, 25, 50, 100, 250, 500
(mA), 1 (A) (with direct current input)
<Voltage> = 25, 50, 100, 250, 500 (mV), 1, 2.5,
5 (V) (with external current sensor input)
Example
:INPUT:CURRENT:RANGE:SIGMA 30A
Description •
[:INPut]:CURRent:RANGe:SIGMA
is valid
only on models with 2 to 4 elements.
•
[:INPut]:CURRent:RANGe:SIGMB
is valid
only on models with 4 elements.
• This command is invalid, if the wiring unit
{ΣA|ΣB} does not exist as a result of the wiring
system setting (
[:INPut]:WIRing
).
6.14 INPut Group