[:INPut]:CURRent:AUTO:ELEMent<x>
Function
Turns ON/OFF the current auto range of the
element or queries the current setting.
Syntax
[:INPut]:CURRent:AUTO:
ELEMent<x> {<Boolean>}
[:INPut]:CURRent:AUTO:ELEMent<x>?
<x> = 1 to 4 (element)
Example
:INPUT:CURRENT:AUTO:ELEMENT1 ON
:INPUT:CURRENT:AUTO:ELEMENT1? ->
:INPUT:CURRENT:AUTO:ELEMENT1 1
[:INPut]:CURRent:AUTO:{SIGMA|SIGMB}
Function
Collectively turns ON/OFF the current auto range
of all elements belonging to wiring unit {ΣA|ΣB}
.
Syntax
[:INPut]:CURRent:AUTO:{SIGMA|
SIGMB} {<Boolean>}
Example
:INPUT:CURRENT:AUTO:SIGMA ON
Description •
[:INPut]:CURRent:AUTO:SIGMA
is valid
only on models with 2 to 4 elements.
•
[:INPut]:CURRent:AUTO:SIGMB
is valid
only on models with 4 elements.
• This command is invalid, if the wiring unit
{SA|SB} does not exist as a result of the wiring
system setting (
[:INPut]:WIRing
).
[:INPut]:CURRent:MODE?
Function
Queries the current mode of all elements.
Syntax
[:INPut]:CURRent:MODE?
Example
:INPUT:CURRENT:MODE? ->
:INPUT:CURRENT:MODE:ELEMENT1 RMS;
ELEMENT2 RMS;ELEMENT3 RMS;
ELEMENT4 RMS
[:INPut]:CURRent:MODE[:ALL]
Function
Collectively sets the current mode of all elements.
Syntax
[:INPut]:CURRent:MODE[:ALL] {RMS|
MEAN|DC|RMEAN}
Example
:INPUT:CURRENT:MODE:ALL RMS
[:INPut]:CURRent:MODE:ELEMent<x>
Function
Sets the current mode of the element or queries
the current setting.
Syntax
[:INPut]:CURRent:MODE:
ELEMent<x> {RMS|MEAN|DC|RMEAN}
[:INPut]:CURRent:MODE:ELEMent<x>?
<x> = 1 to 4 (element)
Example
:INPUT:CURRENT:MODE:ELEMENT1 RMS
:INPUT:CURRENT:MODE:ELEMENT1? ->
:INPUT:CURRENT:MODE:ELEMENT1 RMS
[:INPut]:CURRent:MODE:{SIGMA|SIGMB}
Function
Collectively sets the current mode of all elements
belonging to wiring unit {ΣA|ΣB}.
Syntax
[:INPut]:CURRent:MODE:{SIGMA|
SIGMB} {RMS|MEAN|DC|RMEAN}
Example
:INPUT:CURRENT:MODE:SIGMA RMS
Description •
[:INPut]:CURRent:MODE:SIGMA
is valid
only on models with 2 to 4 elements.
•
[:INPut]:CURRent:MODE:SIGMB
is valid
only on models with 4 elements.
• This command is invalid, if the wiring unit
{ΣA|ΣB} does not exist as a result of the wiring
system setting (
[:INPut]:WIRing
).
[:INPut]:CURRent:RANGe?
Function
Queries the current ranges of all elements.
Syntax
[:INPut]:CURRent:RANGe?
Example
:INPUT:CURRENT:RANGE? -> :INPUT:
CURRENT:RANGE:ELEMENT1 30.0E+00;
ELEMENT2 30.0E+00;
ELEMENT3 30.0E+00;
ELEMENT4 30.0E+00
[:INPut]:CURRent:RANGe[:ALL]
Function
Collectively sets the current ranges of all
elements.
Syntax
[:INPut]:CURRent:RANGe[:ALL]
{<Current>|(EXTernal,<Voltage>)}
When all the input elements of this instrument
are 30 A input elements
• When the crest factor is set to 3
<Current> = 500 (mA), 1, 2, 5, 10, 20, 30 (A) (for
direct current input)
<Voltage> = 50, 100, 200, 500 (mV), 1, 2, 5, 10
(V) (for external current sensor input)
• When the crest factor is set to 6
<Current> = 250, 500 (mA), 1, 2.5, 5, 10, 15 (A)
(for direct current input)
<Voltage> = 25, 50, 100, 250, 500 (mV), 1, 2.5,
5 (V) (for external current sensor input)
When all the input elements of this instrument
are 2 A input elements
• When crest factor is set to 3
<Current> = 5, 10, 20, 50, 100, 200, 500 (mA),
1, 2, (A) (with direct current input)
<Voltage> = 50, 100, 200, 500 (mV), 1, 2, 5, 10
(V) (with external current sensor input)
• When crest factor is set to 6
<Current> = 2.5, 5, 10, 25, 50, 100, 250, 500
(mA), 1 (A) (with direct current input)
<Voltage> = 25, 50, 100, 250, 500 (mV), 1, 2.5,
5 (V) (with external current sensor input)
Example
:INPUT:CURRENT:RANGE:ALL 30A
:INPUT:CURRENT:RANGE:
ALL EXTERNAL,10V
Description For models that have both 2 A input elements and
30 A input elements installed, the direct current
input range cannot be set collectively. Error 863
will occur.