9-5
IM WT3001E-51EN
IEC Harmonic Measurement (Option)
1
2
3
4
5
6
7
8
9
10
11
12
13
Index
9.2 Various Settings Related to IEC Harmonic
Measurement
To set the IEC harmonic measurement, use the Harmonic/Flicker Measurement Software
or set this instrument according to the procedures in the respective section below before
switching this instrument to remote mode using the Harmonic/Flicker Measurement
Software. The underlined sections are settings specific to the IEC harmonic
measurement.
• Selecting the IEC harmonic measurement mode
Section 9.3
If you use the Harmonic/Flicker Measurement Software, this instrument automatically
switches to IEC harmonic measurement mode.
• Changing the number of displayed items of numeric data and Section 7.3
scrolling the display
• Changing the displayed items of numeric data
Section 7.4
• Selecting the measured source
Section 9.4
• Selecting the PLL Source
Section 7.5
• Setting the measured order
Section 7.6
• The maximum order in IEC harmonic measurement mode is 50. Orders exceeding
50 cannot be measured even if you specify such orders.
• If you use the Harmonic/Flicker Measurement Software, the maximum order is
automatically set to 40.
• Selecting the distortion factor equation
Section 7.7
• Selecting the voltage/current grouping
Section 9.5
Note
Do not release the remote mode of this instrument and change the settings while operating this
instrument with the Harmonic/Flicker Measurement Software. Doing so may impede harmonic
measurements conforming to the IEC standard. If you release the remote mode of this
instrument, close the Harmonic/Flicker Measurement Software.