Measurement Range «For procedures, see section 4.3.»
Set the measurement range using an rms level. When directly applying voltage or
current signals to the input element, two types of measurement ranges is available, fixed
range and auto range. When waveforms are displayed, the vertical display range
corresponds to 3 or 6 times the measurement range when the crest factor (see section
4.6) is set to 3 or 6, respectively. For details on waveform display, see section 2.7,
“Waveform Display.”
• Fixed Range
Select each range from a number of choices. The selected range does not switch
even if the amplitude of the input signal changes. For voltage, the maximum and
minimum selectable ranges are 1000 V and 15 V, respectively, when the crest factor
is set to 3. When the crest factor is set to 6, the maximum and minimum selectable
ranges are 500 V and 7.5 V, respectively.
• Auto Range
The measurement range switches automatically depending on the amplitude of the
input signal. The different ranges used in the auto range are the same as those
available for fixed range.
• Range Increase
• When the data of measurement function Urms or Irms exceeds 110% of the
current measurement range, the measurement range is increased.
• When the peak value of the input signal exceeds 330% or 660% of the current
measurement range when the crest factor is set to 3 or 6, respectively, the
range is increased.
• Range Decrease
When the data of the measurement function Urms or Irms is less than or equal to
30% of the measurement range and Upk and Ipk is less than or equal to 300% or
600% of the next lower range when the crest factor is set to 3 or 6, respectively,
the range is decreased.
• Power Range
The measurement ranges (power ranges) of active power, apparent power, and
reactive power are determined by the wiring method, voltage range, and current range
as shown below. For the actual values of the power range, see section 4.3, “Setting
the Measurement Range during Direct Input.”
Wiring System
Power Range
1P2W (single-phase, two-wire)
voltage range
×
current range
1P3W (single-phase, three-wire)
voltage range
×
current range
×
2
3P3W (three-phase, three-wire)
(when the voltage and current ranges on the
corresponding elements are set to the same range)
3P3W (3V3A, three-voltage, three-current)
3P4W (three-phase, four-wire)
voltage range
×
current range
×
3
(when the voltage and current ranges on the
corresponding elements are set to the same range)
2.3 Measurement Conditions