Troubleshooting, Maintenance, and Inspection
11.1 Troubleshooting
Troubleshooting
• If a message is displayed on the screen, read the following pages.
• If servicing is necessary, or if the instrument is not operating correctly after performing
the corrective actions, contact your nearest YOKOGAWA dealer.
Problem and Corrective Action
Reference Section
Nothing shows up on the display when the power switch is turned ON.
Securely connect the power cord to the power connector of the instrument
3.3
and the power outlet.
Set supply voltage within the permitted range.
3.3
Check the display settings.
10.6
Check that the fuse is not blown.
11.5
The displayed data is not correct.
Check that the ambient temperature and humidity are within the
3.2
specifications.
Check that noise is not affecting the measurement.
3.1, 3.5
Check the wiring of the measurement cable.
3.8 to 3.11
Check the wiring system.
3.8 to 3.11, 4.1
Check that the line filter is turned OFF.
4.8
Check the measurement period settings.
4.7
Check the FAQ at the following URL.
–
http://www.yokogawa.com/tm/
Power cycle the instrument.
3.4
Keys do not work.
Check the REMOTE indicator. If the REMOTE indicator is ON, press
–
the LOCAL key to turn it OFF.
Check that key lock is OFF.
10.7
Perform a key test. If it fails, servicing is required.
11.3
Triggering does not work.
Check the trigger conditions.
6.4
Check that the trigger source is being applied.
6.4
Unable to make harmonic measurements.
Check the PLL source settings.
–
*1
Check that the PLL source input signal is within the specifications.
–
*1
Unable to print on the built-in printer.
The printer head may be damaged or worn out.
–
Servicing required.
Unable to recognize the medium.
Check the format of the medium. If necessary, format the medium.
9.2
The medium may be damaged.
–
Unable to save data to the selected medium.
If necessary, format the storage medium.
9.2
Check the free space on the medium. Remove unnecessary files as
9.2
necessary or use a new medium.
Unable to set or control the instrument via the communication interface.
Check that the GP-IB address or the parameter settings of the serial
–
*2
interface satisfy the specifications.
Check that the interface satisfies the electrical and mechanical
–
*2
specifications.
*1 See the
Expansion Function User’s Manual IM760301-51E
.
*2 See the
Communication Interface User’s Manual IM760301-17E
.