Power Measurement
5.2
Selecting the Voltage/Current Mode (RMS,
MEAN, DC, or RMEAN)
Procedure
START
STOP
REMOTE
CAL
SENSOR RATIO
MEASURING
RMS
MEAN
DC
RMEAN
CURRENT RANGE
VOLTAGE RANGE
MEASURING
RMS
MEAN
DC
RMEAN
ALL
INPUT INFO.
LOWER ITEM
LOWER FORM
ALL
USER SET
DISPLAY
ITEM & ELEMENT
2
3
4
COMPEN
ELEMENT
ELEMENT
1
WIRING
MODE
AUTO
MODE
AUTO
EXT SENSOR
NUMERIC
WAVE
OTHERS
ITEM
FORM
U / I / P
WP/q/
TIME
USER
ELEMENT
UPDATE
RATE
HOLD
SINGLE
INTEG
LOCAL
SHIFT
RESET
SET
PAGE
PAGE
Cursor keys
To exit the menu during
operation, press
ESC
located above the soft
keys.
In the procedural
explanation below, the
phrase “press the
cursor
keys
” may be used. This
phrase refers to the
procedures for selecting
items and entering values
and character strings. For
details on the procedures,
see section 3.14.
• Selecting the Target Element
1.
Press
ELEMENT
to select the target element. The corresponding element
indicator above the key illuminates.
• Selecting the Voltage Mode
2.
Press
VOLTAGE RANGE MODE
to select one of the modes from RMS to RMEAN.
3.
Repeat steps 1 and 2 to select the voltage mode of all elements.
• Selecting the Current Mode
4.
Press
CURRENT RANGE MODE
to select one of the modes from RMS to
RMEAN.
5.
Repeat steps 1 to 4 to select the current mode of all elements.
Explanation
• The voltage and current modes can be selected from the following four modes.
• RMS:
True rms value
• MEAN:
Rectified mean value calibrated to the rms value
• DC:
Simple average
• RMEAN:
Rectified mean value
• The initial setting is RMS. For the method of computing the voltage and current in
each mode, see appendix 1.
• When the independent setting of input elements is OFF, the element switches
according to the wiring system.
• To measure the RMS and MEAN of voltage simultaneously, use the user-defined
function. For details, see section 5.4.
• The voltage and current mode selections are invalid in wide bandwidth harmonic
measurement mode and IEC harmonic measurement mode. In this case, the total
value of the rms values of each order in the range between the maximum and
minimum values of the measured order specified in the harmonic measurement
(option) menu is displayed.
• The voltage and current mode selections are invalid in voltage fluctuation and flicker
measurement mode.
Содержание wt3000
Страница 1: ...Precision Power Analyzer WT3000 IM 760301 01E 8th Edition WT3000 Manual Vol 1 3 ...
Страница 3: ......