Power Measurement
• Selecting the Number of Displayed Items
Select the number of numeric data items that are displayed simultaneously from the
choices below.
• 4 Items
• When the display mode is Numeric, N Wave, N Bar
*1
, and
N Trend, four numeric data values are displayed in one column.
• 8 Items
• When the display mode is Numeric, eight numeric data values are displayed in
one column.
• When the display mode is N Wave, N Bar
*1
, and N
Trend, eight numeric data values are displayed in two columns.
• 16 Items
• When the display mode is Numeric, N Wave, N Bar
*1
, 16
numeric data values are displayed in two columns.
• All Items
A table is displayed indicating the numeric data of items with measurement
functions listed vertically and symbols indicating elements and wiring units listed
horizontally. The number of displayed items varies depending on the number of
installed elements. The displayed item cannot be changed independently using
the ITEM key. Switch the display by page scrolling.
• Single List
*1
• When the display mode is Numeric, 42 numeric data values for a single
measurement function are displayed in two columns.
• When the display mode is N Wave, N Bar
*1
, 22 numeric data
values for a single measurement function are displayed in two columns.
• Dual List
*1
• When the display mode is Numeric, 22 numeric data values each for two
measurement functions are displayed in each column.
• When the display mode is N Wave, N Bar
*1
, 11 numeric data
values each for two measurement functions are displayed in each column.
*1 Displayed only on models with the advanced computation (/G6) or harmonic
measurement (/G5) option. For the procedure of changing the displayed items on the
Single List or Dual List, see chapter 7 in the Expansion Function User’s Manual
IM760301-51E.
Note
•
[-------] (no data) is displayed in places where the measurement function is not selected or
where no numeric data is present.
•
If Urms, Umn, Udc, Urmn, Irms, Imn, Idc, or Irmn exceeds 140% of the measurement range,
overrange [-OL-] is displayed.
•
P shows overrange [-OL-] if the measured values of either the voltage or current exceeds
140% of the measurement range.
•
If the measured or computed result cannot be displayed using the specified decimal position
or unit, overflow [-OF-] is displayed.
•
If Urms or Irms is less than or equal to 0.3% (when the crest factor is set to 3, less than or
equal to 0.6% if the crest factor is set to 6) or Umn, Urmn, Imn, or Irmn is less than or equal to
1% of the measurement range (when the crest factor is set to 3, less than or equal to 2% if
the crest factor is set to 6), Urms, Umn, Urmn, Irms, Imn, Irmn, and the measurement
functions that are determined using these measurement functions display zeroes.
λ
or
φ
displays an error.
•
The value of U+pk and I+pk when there are no positive measured values, and the value of U-
pk and I-pk when there are no negative measured values are as follows.
•
Products of Firmware Version 4.04 or Later: The maximum value measured is displayed
for U+pk and I+pk, and the minimum value measured is displayed for U-pk and I-pk.
•
Products of Firmware Versions Prior to 4.04: Zero is displayed for U+pk, I+pk, U-pk and I-pk.
•
If the measured value of frequency is outside the measurement range, fU or fI displays an error.
•
If the power factor
λ
is greater than 1 and less than or equal to 2,
λ
becomes [1].
φ
displays zero.
•
If the power factor
λ
is greater than 2,
λ
and
φ
display errors.
•
If the voltage or current mode is not RMS, CfU or CfI displays no numeric data [-------].
5.1 Displaying Numeric Data and Changing the Displayed Items