![YOKOGAWA WT300 Скачать руководство пользователя страница 121](http://html.mh-extra.com/html/yokogawa/wt300/wt300_user-manual_911408121.webp)
App-1
IM WT310-02EN
Appendix
1
2
3
4
5
6
7
App
Index
Appendix 1 How to Make Accurate Measurements
Effects of Power Loss
By wiring a circuit to match the load, you can minimize the effects of power loss on measurement
accuracy. We will discuss the wiring of the DC power supply (SOURCE) and a load resistance (LOAD)
below.
When the Measured Current Is Relatively Large
Connect the voltage measurement circuit between the current measurement circuit and the load.
The current measurement circuit measures the sum of i
L
and i
V
. i
L
is the current flowing through the
load of the circuit under measurement, and i
V
is the current flowing through the voltage measurement
circuit. Because the current flowing through the circuit under measurement is i
L
, only i
V
reduces
measurement accuracy. The input resistance of the voltage measurement circuit of the WT300 series
is approximately 2 MΩ. If the input voltage is 600 V, i
V
is approximately 0.3 mA (600 V/2 MΩ). If the
load current i
L
is 3 A or more (the load resistance is 200 Ω or less), the effect of i
V
on the measurement
accuracy is 0.01% or less. If the input voltage is 100 V and the current is 5 A, i
V
= 0.05 mA (100 V/2
MΩ), so the effect of i
V
on the measurement accuracy is 0.001% (0.05 mA/5 A).
SOURCE
LOAD
WT300 series
V
C
i
V
i
L
LOAD
SOURCE
U
I
C
V
±
±
±
±
As a reference, the relationships between the voltages and currents that produce effects of 0.01%,
0.001%, and 0.0001% are shown in the figure below.
600
400
200
0
Measured current (A)
Measured voltage (V)
0.01% effect
0 5 10 15 20 25 30 35 40
0.001% effect
0.0001% effect
Smaller effect
3
When the Measured Current Is Relatively Small
Connect the current measurement circuit between the voltage measurement circuit and the load.
In this case, the voltage measurement circuit measures the sum of e
L
and e
I
. e
L
is the load voltage,
and e
I
is the voltage drop across the current measurement circuit. Only e
I
reduces measurement
accuracy. For example, the input resistance of the current measurement circuit of the WT332/WT333
is approximately 6 mΩ. If the load resistance is 600 Ω, the effect on the measurement accuracy is
approximately 0.001% (6 mΩ/600 Ω). For the input resistances of the WT310 and WT310HC, see
chapter 7.
SOURCE
LOAD
V
C
WT300 series
e
L
e
I
SOURCE
LOAD
U
I
C
V
±
±
±
±
Appendix