2-4
IM 760401-01E
2.3
Auto Range Monitor, Overrange, and Error
Displays during Measurement
Overrange Display
During normal measurement, overrange occurs when the measured voltage V or
measured current A exceeds 140% of the rated range. When using auto range, the
range automatically steps up. Overrange occurs when 140% of the maximum range
is exceeded. Overrange is indicated as follows: For the conditions in which overrange
display occurs in harmonic measurement, see section 7.1.
Computation Over Display
If the computed value overflows during the computation process, it is indicated as
follows:
Auto Range Monitor
The indicator illuminates when the input signal meets the conditions for auto range
switching. For details on the auto range function, see section 4.4.
CHECK RANGE
VOLTAGE
CURRENT
CHECK RANGE
VOLTAGE CURRENT
WT210 indicator
WT230 indicator
Display When the Measured Value is too Small
If the measured voltage or current drops below 0.5% (less than or equal to 1% if the
crest factor is set to 6) of the rated range, the condition is indicated as shown below.
This applies only when the measurement mode (section 4.1, “Selecting the
Measurement Mode”) is set to RMS or VOLTAGE MEAN.
Function
Display
V (voltage)
A (current)
Displays zero.
VA (apparent power)
var (reactive power)
PF (power factor)
deg (phase angle)
Measurement Abort/No Data Display (Bar Display)
If you change the displayed information by changing the measurement range or the
function/element or there is no data to be displayed, the following bar display appears
temporarily.
Содержание WT210
Страница 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Страница 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Страница 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Страница 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Страница 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...