15-5
IM 760401-01E
Maintenance and Troubleshooting
15
9.
Repeat steps 2 to 6 to adjust channel 2.
10.
Repeat steps 2 to 6 for all channels to be adjusted.
11.
Press The SHIFT key and display C will change to “dA”. This completes the D/A
output adjustments. When you press the RESET key instead of the SHIFT key,
the carried out adjustments will become invalid.
After Finishing Adjustments
After having finished all adjustments, turn the power OFF and ON again.
Communication Commands for Adjustment
Command Description
CAL1
Enters range adjustment mode.
CR0
Switches to 15 V range in range adjustment mode.
CR1
Switches to 30 V range in range adjustment mode.
CR2
Switches to 60 V range in range adjustment mode.
CR3
Switches to 150 V range in range adjustment mode.
CR4
Switches to 100 mA range in range adjustment mode (WT210only).
CR5
Switches to 500 mA range in range adjustment mode.
CR6
Switches to 1 A range in range adjustment mode.
CR7
Switches to 2 A range in range adjustment mode.
CR8
Switches to 5 A range in range adjustment mode.
CR11
Switches to 150 V range in range adjustment mode with the line filter turned ON.
CR12
Switches to 1 A range in range adjustment mode with the line filter turned ON.
CAN
Exits range adjustment mode discarding the adjustments. Corresponds to the
RESET key.
END
Exits range adjustment mode keeping the adjustments. Corresponds to the SHIFT
key.
ENT
Confirms the adjustment value. Corresponds to the ENTER key.
CAL2
Enters external sensor input range adjustment mode.
EX0
No external sensor input option.
EX1
Option /EX1 (2.5 V, 5 V, and 10 V ranges)
EX2
Option /EX2 (50 mV, 100 mV, and 200 mV ranges)
CR9
/EX2 Switch to 200 mV range.
CR10
/EX1 Switch to 10 V range.
CAN
Exits range adjustment mode discarding the adjustments. Corresponds to the
RESET key.
END
Exits range adjustment mode keeping the adjustments. Corresponds to the SHIFT
key.
ENT
Confirms the adjustment value. Corresponds to the ENTER key.
CAL3
Enter D/A output adjustment mode.
CHm
Switches channels in D/A output adjustment mode.
m = 1 to 12
CDm,n
Enter the actual output value in D/A output adjustment mode.
m = 1 to 12, n = actual output value.
DO0
E5 V output in D/A output adjustment mode.
DO1
Execute –5 V output in D/A output adjustment mode.
CAN
Exits range adjustment mode discarding the adjustments. Corresponds to the
RESET key.
END
Exits range adjustment mode keeping the adjustments. Corresponds to the SHIFT
key.
ENT
Confirms the adjustment value. Corresponds to the ENTER key.
CAL0
Exits adjustment (returns to normal measurement mode).
OD
Communication output request for measured/computed data. The output format is
normally default.
Note
•
In case of D/A adjustment, change the channel using the CHm command, then carry out DO0
or DO1 command, and the request output using the CDm,n command. After the CDm,n
command is being executed by +5 V or –5 V, make sure to confirm by the ENT command.
•
After the display has been stabilized in the range adjustment mode or external sensor input
range adjustment mode, execute the ENT command.
15.1 Adjustments
Содержание WT210
Страница 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Страница 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Страница 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Страница 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Страница 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...