IM 760401-01E
Communication Commands 2 (System of Commands Complying to the IEEE 488.2-1992 Standard)
14
Output Example of Normal Measurement Data
• Output example for model 760502 after having sent the following commands.
(Sent)
MEASURE:NORMAL:ITEM:PRESET NORMAL
MEASURE:NORMAL:VALUE?
(Received data)
10.004E+00,10.002E+00,10.003E+00,49.041E+00,49.052E+00,
49.047E+00,429.00E+00,429.02E+00,0.8580E+03
(Data contents)
V1:
10.004E+00
V3:
10.002E+00
V
∑
:
10.003E+00
A1:
49.041E+00
A3:
49.052E+00
A
∑
:
49.047E+00
W1:
429.00E+00
W3:
429.02E+00
W
∑
:
0.8580E+03
• Output example for model 760503 where measurement data first have been
stored during integration, and while recalling these data, the following
commands have been sent.
(Sent)
MEASURE:NORMAL:ITEM:PRESET INTEGRATE
MEASURE:NORMAL:VALUE?
(Received data)
10,428.60E+00,428.10E+00,428.80E+00,
1.2855E+03,71.450E+00,71.370E+00,71.490E+00,214.31E+00,
8.2342E+00,8.2354E+00,8.2519E+00,24.721E+00,0,10,0
(Data contents)
Recalled data number: 10
W1:
428.60E+00
W2:
428.10E+00
W3:
428.80E+00
W
∑
:
1.2855E+03
WH1:
71.450E+00
WH2:
71.370E+00
WH3:
71.490E+00
WH
∑
:214.31E+00
AH1:
8.2342E+00
AH2:
8.2354E+00
AH3:
8.2519E+00
AH
∑
:
24.721E+00
Elapsed integration time: 0 (hours), 10 (minutes), 0 (seconds)
Data Format of Harmonic Measurement
All data will be output in the <NR3> format. (mantissa: max. 5 exponent: 2
digits)
Output Format of Harmonic Measurement
The communication output is set ON by any of the commands starting with
“
MEASure:HARMonics:ITEM
” and the harmonic measurement data or frequency of PLL
source (
SYNChronize
) are output according to the following order of priority. Besides,
in case of recalling normal measurement or integration data, the data number will be
output in <NR1> format as well.
(
0.
Data number in case of recalling)
1.
Frequency of PLL source (
SYNChronize
)
2.VTHD
3.V
4.VCON
5.ATHD
6.A
7.ACON
8.PF
9.W
10.WCON
11.VDEG
12.ADEG
Harmonic measurement data will be output for all applicable elements. To find out to
which element the data correspond, use the
HARMonics:ELEMent?
command.
• Frequency of PLL Source (
SYNChronize
) : 1 data
Outputs the fundamental frequency (
VHz/AHz
) of the voltage/current for which the
PLL source has been set. The input of the PLL source can be found out using
HARMonics:SYNChronize?
.
•
VTHD,ATHD
: 1 data
Outputs the harmonic distortion factor of voltage/current. (for either iEC or CSA). The
used computation method can be found out using the
HARMonics:THD?
command.
•
V,A,W
: 51(or 31) data
Rms values of the 1st to 50(or 30)th order
→
fundamental measured value (1st
order)
→
harmonic measurement value (2nd order)
→
· · ·
→
harmonic measurement
value (50(or 30)th order)
14.3 Commands
Содержание WT210
Страница 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Страница 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Страница 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Страница 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Страница 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...