14-32
IM 760401-01E
Data Format/Output Format of Normal and Harmonic Measurement Data <ASCII>
The data format/output format of normal and harmonic measurement data which is
requested by
MEASure[:NORMal]:VALue?
or
MEASure:HARMonics:VALue?
, is as follows.
Data Format of Normal Measurement Data
• All data of the <harmonic measurement function> are output in the <NR3> format.
(Example) 123.45E+00
V,A,W,VA,VAR,PF,DEGR,VHZ,AHZ,VPK,APK,MATH
→
mantissa: max. 5
exponent: 2 digits
WH,WHP,WHM,AH,AHP,AHM
→
mantissa: max. 6 exponent: 2 digits
(max. 5 digits in case of negative value)
• The sign of the mantissa will only be applied in case of negative values. However,
phase lead and lag (in case of phase angle (DEG)) will be shown as follows.
LEAD
→
+180.0E+00
LAG
→
-180.0E+00
in phase
→
0.0E+00
(The mantissa will be proceeded by a space)
• In case of overrange or computation over,
“9.9E+37“(+
∞
)
will be output.
(i.e. in case the display shows –oL–, –oF–, PFErr, dEGEr, ErrLo, or ErrHi)
• In case no data is present (i.e. the display shows – – – – –),
“9.91E+37“ (NAN)
will be
output.
• The elasped integration time is output as hours, minutes, seconds in the <NR1>
format.
(Example) 999,59,59
• Only one of the six parameters, VHZ1, VHZ2, VHZ3, AHZ1, AHZ2, or AHZ3, can be
measured for frequency. Data that is not measured is set to no data
(9.91E+37)
.
• VHZ
Σ
, AHZ
Σ
, VPK
Σ
, and APK
Σ
always output no data
(9.91E+37)
.
Output Format of Normal Measurement Data
The communication output is set ON by any of the commands starting with
“
MEASure[:NORMal]:ITEM
” and the normal measured/computed data or elapsed
integration time are output according to the following order of priority. Besides, in
case of recalling normal measurement or integration data, the data number will be
output in <NR1> format as well. Data will be output in the following order
corresponding to each element. However, note that for model 760401 only element 1
is valid, and for model 760502 only element 1, 3 and S are valid.
(
0.
Data number in case of recalling)
1. V1
→
V2
‘
→
V3
→
V
∑
2. A1
→
A2
→
A3
→
A
∑
3. W1
→
W2
→
W3
→
W
∑
4. VA1
→
VA2
→
VA3
→
VA
∑
5. VAR1
→
VAR2
→
VAR3
→
VAR
∑
6. PF1
→
PF2
→
PF3
→
PF
∑
7. DEGR1
→
DEGR2
→
DEGR3
→
DEGR
∑
8. VHZ1
→
VHZ2
→
VHZ3
→
VHZ
∑
9. AHZ1
→
AHZ2
→
AHZ3
→
AHZ
∑
10. WH1
→
WH2
→
WH3
→
WH
∑
11. WHP1
→
WHP2
→
WHP3
→
WHP
∑
12. WHM1
→
WHM2
→
WHM3
→
WHM
∑
13. AH1
→
AH2
→
AH3
→
AH
∑
14. AHP1
vAHP2
→
AHP3
→
AHP
∑
14. AHM1
→
AHM2
→
AHM3
→
AHM
∑
16. TIME
(elapsed integration time)
Each data is divided by a comma”
,
” and is ended by the terminator <RMT>.
14.3 Commands
Содержание WT210
Страница 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Страница 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Страница 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Страница 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Страница 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...