14-22
IM 760401-01E
14.3.5 DISPlay Group
The commands in the DISPlay group are used to make settings relating to, and inquiries about display. This allows
you to make the same settings and queries as when using the FUNCTION key or ELEMENT key on the front panel.
:DISPlay
<x>
:
MODE
<Space>
VALue
RANGe
ESCaling
?
FUNCtion
<Space>
?
ELEMent
<Space>
<NRf>
SIGMa
?
<Display function>
?
;
RESolution
<Space>
HIGH
LOW
?
Example
[CONFIGURE]:VOLTAGE:RANGE 600V
[CONFIGURE]:VOLTAGE:RANGE?
-> :CONFIGURE:VOLTAGE:RANGE 600.0E+00
14.3 Commands
DISPlay<x>?
Function
Queries all the display settings.
Syntax
DISPlay<x>?
<x>= 1 to 3
1:Display A
2:Display B
3:Display C
Example
DISPlay1? -> :DISPLAY1:MODE VALUE;
FUNCTION V;ELEMENT 1;RESOLUTION HIGH
[CONFigure]:WIRing
Function
Sets the wiring method/queries the current
setting.
Syntax
[CONFigure]:WIRing {P1W2|P1W3|P3W3|
P3W4|V3A3}
[CONFigure]:WIRing?
Example
[CONFIGURE]:WIRING P1W3
[CONFIGURE]:WIRING?
-> :CONFIGURE:WIRING P1W3
Description The selections stand for the following.
P1W3: Single-phase, three-wires (only for
WT230)
P3W3: Three-phase, three-wires (only for
WT230)
P3W4: Three-phase, four-wires (only for WT230
3-phase, 4-wire model)
V3A3: Three-voltage, three-current (only for
WT230 3-phase, 4-wire model)
DISPlay<x>:ELEMent
Function
Sets the element to be displayed/queries the
current setting.
Syntax
DISPlay<x>:ELEMent {<NRf>|SIGMa}
DISPlay<x>:ELEMent?
<x>= 1 to 3
1:Display A
2:Display B
3:Display C
{<NRf>}=1
(WT210 single-phase
model)
1, 3
(WT230 three-phase,
three-wire model)
1 to 3 (WT230 three-phase,
four-wire model)
Example
DISPLAY1:ELEMENT 1
DISPLAY1:ELEMENT? -> :DISPLAY1:
ELEMENT 1
Содержание WT210
Страница 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Страница 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Страница 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Страница 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Страница 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...