14-4
IM 760401-01E
Program message unit format
The format of a program message unit is shown below.
,
<Program header>
<Program data>
Space
<Program header>
A program header is used to indicate the command
type. For details, refer to page 14-5.
<Program data>
If certain conditions are required for the execution of a
command, program data must be added. Program
data must be separated from the header by a space
(ASCII code “20H”). If multiple items of program data
are included, they must be separated by a “
,
”
(comma).
Header
Data
Example
:CONFIGURE:AVERAGING:TYPE LINEAR,8<PMT>
Response Messages
The data returned by this instrument to the controller is
called a response message. The format of a response
message is shown below.
<RMT>
;
<Response message unit>
<Response message units>
A response message consists of one or more
response message units: each response message unit
corresponds to one response.
Response message units are delimited by a “
;
”.
For the response message format, refer to the next
item.
Unit
Example
:CONFIGURE:VOLTAGE:RANGE 15.0E+00;AUTO 0<RMT>
Unit
<RMT>
RMT is the terminator used for every response
message. Only one type of response message is
available; NL^END.
Response message unit format
The format of a program message unit is shown below.
,
<Rsps. header>
<Response data>
Space
<Response header>
A response header sometimes precedes the response
data. Response data must be separated from the
header by a space. For details, refer to page 14-7.
<Response data>
Response data is used to define a response. If
multiple items of response data are used, they must be
separated by a “
,
” (comma).
Header
Data
Data
Example
:500.0E-03<RMT> :CONFIGURE:MODE RMS<RMT>
If a program message contains more than one query,
responses are made in the same order as the queries.
Normally, each query returns only one response
message unit, but there are some queries which return
more than one response message unit. The first
response message unit always responds to the first
query, but it is not always true that the ‘n‘th unit always
responds to the ‘n‘th query. Therefore, if you want to
make sure that a response is made to each query, the
program message must be divided up into individual
messages.
Points to Note concerning Message Transmission
• It is always possible to send a program message if
the previous message which was sent did not
contain any queries.
• If the previous message contained a query, it is not
possible to send another program message until a
response message has been received. An error will
occur if a program message is sent before a
response message has been received in its entirety.
A response message which has not been received
will be discarded.
• If an attempt is made by the controller to receive a
response message, even if there it no response
message, an error will occur. An error will also
occur if the controller makes an attempt to receive a
response message before transmission of a
program message has been completed.
• If a program message of more than one unit is sent
and some of the units are incomplete, this
instrument receives program message units which
the instrument thinks complete and attempts to
execute them. However, these attempts may not
always be successful and a response may not
always be returned, even if the program message
contains queries.
14.2 Program Format
Содержание WT210
Страница 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Страница 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Страница 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Страница 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Страница 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...