Output Format
The output format depends on the selected output items which can be selected by the
“OH” command.
• In case of voltage and current
All computed values of
the 1st to 50th order
,
Terminator
Measured value for
fundamental (1st order)
,
Terminator
END
Terminator
Line 1
Line 2
Line 3
Line 51
Line 52
Measured value for 50th
harmonic
,
Terminator
Measured value for 2nd
harmonic
,
Terminator
Frequency
Relative harmonic content
for 2nd harmonic
Relative harmonic content
for 50th harmonic
harmonic distortion
• In case of active power
All computed values of
the 1st to 50th order
,
Terminator
Measured value for
fundamental (1st order)
,
Terminator
END
Terminator
Line 1
Line 2
Line 3
Line 51
Line 52
Measured value for
50th harmonic
,
Terminator
Measured value for
2nd harmonic
,
Terminator
Frequency
Relative harmonic content
for 2nd harmonic
Relative harmonic content
for 50th harmonic
Power factor
• In case of phase angle
,
Terminator
END
Terminator
Line 1
Line 2
Line 3
Line 50
Line 51
Phase angle between
fundamentals
of voltage and current
Frequency
Phase angle between
fundamental and
2nd harmonic of current
,
Terminator
Phase angle between
fundamental and 2nd
harmonic of voltage
,
Terminator
Phase angle between
fundamental and
3rd harmonic of voltage
,
Terminator
Phase angle between
fundamental and
50th harmonic of voltage
Phase angle between
fundamental and
3rd harmonic of current
Phase angle between
fundamental and
50th harmonic of current
• In case of ALL setting
The data will be output in the sequence voltage
→
current
→
active power
→
phase
angle
→
END <terminator>
• The output format of each item is as described for each item above;
• The END line is not output for each item, but after finishing the entire output
operation.
Output Format of Setup Parameters and Error Codes
See the response examples of “OS” and “OE” commands in section 13.1. For a
description of the displayed information of the response examples, see the explanation
of the respective command in section 13.1.
10.4 Output Measured/Computed Data, Setup Parameters, and Error Codes
Содержание WT210
Страница 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Страница 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Страница 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Страница 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Страница 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...