15.4 Error Codes and Corrective Actions
Error Codes for Operation and Measurement
Error Code
Description
Corrective Action
Reference page
11
Received a command not used by this instrument.
Check for error in the command sent.
10-7
12
Parameter value specified is outside the allowed
Correct the value.
Ch. 13,Ch. 14
range.
13
Attempted to execute a key operation or
Check whether integration is in
6-4, 10-7
received a communications command, while
progress or is interrupted.
integration was running or was interrupted, that
cannot be executed or received in such a state.
14
Attempted to set auto range while external sensor
It is not possible to set auto
4-4
range is selected.
range while external sensor
range is selected.
15
Attempted to execute a command or key
Check whether the command or
–
operation that was protected.
key operation is correct.
16
Attempted to execute a key operation or
Check whether harmonic
7-3
received a communications command, while
analysis is in progress or is
harmonic analysis was being performed or
interrupted.
was interrupted, that cannot be executed or
received in such a state.
17
Print output time-out.
9-9
18
No data to be printed or not in the printing mode.
9-9
19
Attempted to execute a key operation or
–
received a communications command, while
storing/recalling of data being performed.
30
Invalid file data.
–
31
File is damaged.
–
32
No data stored in the internal memory.
Store data in the internal
Ch. 8
memory or select the proper file
to be stored.
33
No space to store data in the internal memory.
8-1
41
• Attempted to start integration while there is
Reset integration.
6-3
an overflow condition.
• Attempted to start integration after
integration time has reached its preset value.
42
Attempted to start integration while integration is
6-4
in progress.
43
Measurement stopped due to overflow
6-4
during integration or due to a power failure.
44
Attempted to stop integration even though
6-4
integration was not in progress.
45
Attempted to reset integration even though
6-4
integration was not in progress or integration
mode was not selected.
46
Attempted to start integration while
6-4
measurement of peak overflow was in
progress or during an overrange condition.
47
Attempted to start integration in continuous
Set a correct preset time.
6-2
integration mode when integration preset
time was set to “0”.
50
A/D conversion time-out.
–
51
Measurement data overflow occurred. “
” is displayed.
2-3
52
Voltage peak overflow occurred.
2-3
V OVER indicator lights up.
53
Current peak overflow occurred.
2-3
A OVER indicator lights up.
54
Power factor exceeded “2” during measurement of power factor.
5-2
55
“
” was displayed at the end of power factor
5-2, 5-3
computation during measurement of phase angle.
56
Input level was too low or below measurement range
5-4
during measurement of frequency. “
” is displayed.
57
Measured frequency exceeded the measurement range.
5-4
“
” is displayed.
58
Computation overflow occurred. “
” is displayed.
2-3
59
Harmonic analysis becomes “
”.
7-1
390
Overrun error.
Lower the baud rate.
11-3