15-5
Maintenance and Troubleshooting
IM 253421-01E
15
15.2 Calibration
Required Equipment
DC Voltage/Current Standard
recommended: Yokogawa 2552, 2550
AC Voltage/Current Standard
recommended: Yokogawa 2558, 2558-S7
or 9100 (up to 400 Hz)
or Fluke 5200A
+
5215A or 5200A
+
5220A
Digital Power Meter
recommended: Yokogawa WT2000 or 2531
2ch Synchronizer
recommended: Yokogawa FG120
Calibration of DC Voltage, Current and Power
Wiring
Connect the DC voltage and DC current standard as follows.
H
L
H
L
V
±
A
±
H
L
EXT
±
(A)
EUT:Equipment under test
• Direct input
DC voltage
standard
DC Amp.
• External sensor input (equipped with option /EX1)
EUT
DC voltage
standard
H
L
EXT
(A)
• External sensor input (equipped with option /EX2)
EUT
DC voltage
standard
±