11-1
IM 253421-01E
Using the RS-232-C Interface (Optional)
11
11.1 RS-232-C Interface Functions and Specifications
This instrument is equipped with a RS-232-C interface in accordance with your
preference. This interface permits remote control from a controller such as a personal
computer, and output of various data.
Overview of the RS-232-C Interface
The table below shows functions that are available in each mode.
Mode
Function
Normal mode
Reception
Functions performed by key operations (except for
LOCAL key and power ON/OFF)
measured, computed, and analysis data output
request
setting parameter output request
error code output request
Transmission
measured, computed, and analysis data output
setting parameter output
error code output
status byte output
Talk-only mode
Transmission
measured, computed, and analysis data output
Normal Mode
This mode is equivalent to the the addressable mode A of the GP-IB function, and
enables reception of commands and transmission of data. Measured data is output
on reception of the “OD” command.
488.2 Mode
This mode allows receiving of commands conforming to the IEEE St'd 488.2-1987
protocol.
Talk-only Mode
This mode is equivalent to the Talk-only mode of the GP-IB function. Only measured
data can be output and commands cannot be received.
There is no equivalent to the addressable mode B of the GP-IB function.
Print Mode
This mode is used to output harmonic analysis data to an external plotter or printer.
For details, see section 9.9.
Chapter 11 Using the RS-232-C Interface (Optional)