9-19
IM 253421-01E
Using External Input/Output
9
SHIFT
SETUP
OUTPUT
ENTER
3.
Activate the output
(Display C)
2.
1.
• Activating the Output
Select data or setting
parameters
(Display C)
ENTER
5.
End
4.
Explanation
Setting the Output (Printing) Mode
This setting is to select whether you are printing out on a plotter or a printer.
HPGL: For printing on an external, HPGL - compatible plotter.
ESCP: For printing on an external, ESC/P - compatible printer.
PCL: For printing on an external, PCL5 (printer language of HP) - compatible printer.
Setting the Output Contents in case of Normal Measurement
All measured/computed data will be output. All measured and computed data are
output. However, for frequency, if either one of the display functions, V Hz or A Hz, is
turned on, the frequency of the corresponding function is output. If both functions are
turned OFF, the frequency of the function that was lighted last will be output.
Setting the Output Items and the Element in case of Harmonic Analysis
• Setting the Output Item
Set one of the following items. The selected item is printed on the external plotter
or printer. The initial value is V.
• V: Prints the numerical values of the analysis value and relative harmonic
content of the voltage;
• A: Prints the numerical values of the analysis value and relative harmonic
content of the current;
• P: Prints the numerical values of the analysis value and relative harmonic
content of the active power;
• dEG: Prints the numerical values of the phase angle;
• G-V: Prints the numerical values
*
and the graph of the analyzed voltage values;
• G-A: Prints the numerical values
*
and the graph of the analyzed current values;
• G-P: Prints the numerical values
*
and the graph of the analyzed active power
values;
• G-Vd: Prints the numerical values
*
and the graph of the phase angle between
each voltage of the 2nd to 50th order and the fundamental (1st order);
• G-Ad: Prints the numerical values
*
and the graph of the phase angle between
each current of the 2nd to 50th order and the fundamental (1st order);
• CG-V: Prints the numerical values
*
and the graph of the relative harmonic
content of voltage;
• CG-A: Prints the numerical values
*
and the graph of the relative harmonic
content of current;
• CG-P: Prints the numerical values
*
and the graph of the relative harmonic
content of active power;
• ALL: Prints the numerical values
*
and the graph of the analysis values and
relative harmonic content of voltage and current (V and A are both printed).
*
HPGL/PCL plotters print both numerical values and the graph, but ESCP printers only
print the graph.
9.9 Outputting to an External Plotter or Printer